APPENDIX C - VARIABLES IN VISCOSITY MEASUREMENTS

As with any instrument measurement, there are variables that can affect a Rheometer measurement. These variables
may be related to the instrument (Rheometer), or the test fluid. Variables related to the test fluid deal with the
rheological properties of the fluid, while instrument variables would include the Rheometer design and the spindle
geometry system utilized.
RHEOLOGICAL PROPERTIES
Fluids have different rheological characteristics that can be described by Rheometer measurements. We can then
work with these fluids to suit our lab or process conditions.
There are two categories of fluids:
Newtonian
Non-Newtonian
Time Independent
Pseudoplastic
Time Dependent
Thixotropic
AMETEK Brookfield publication, "More Solutions to Sticky Problems", includes a more detailed discussion of
rheological properties and non-Newtonian behavior.
Rheometer related variables
Most fluid viscosities are found to be non-Newtonian. They are dependent on Shear Rate, time of test and the spindle
geometry conditions. The specifications of the Rheometer spindle and chamber geometry will affect the viscosity
readings. If one reading is taken at 2.5 RPM, and a second at 50 RPM, the two cP values produced will be different
because the readings were made at different shear rates. The faster the spindle speed, the higher the shear rate.
The shear rate of a given measurement is determined by: the rotational speed of the spindle, the size and shape of
the spindle, the size and shape of the container used and therefore the distance between the container wall and the
spindle surface.
A repeatable viscosity test should control or specify the following:
Test temperature
Sample container size (or spindle/chamber geometry)
Sample volume
Rheometer model
M19-2101 REVISION 2.0
These fluids have the same viscosity at different Shear Rates (different RPMs) and
are called Newtonian over the Shear Rate range they are measured.
These fluids have different viscosities at different shear rates (different RPMs). They
fall into two groups:
Time Independent non-Newtonian
Time Dependent non-Newtonian
A pseudoplastic material displays a decrease in viscosity with an increase in shear
rate, and is also known as "shear thinning". If you take Rheometer readings from
a low to a high RPM and then back to the low RPM, and the readings fall upon
themselves, the material is time independent pseudoplastic (shear thinning).
A thixotropic material has decreasing viscosity under constant shear rate. If you set
a Rheometer at a constant speed recording cP values over time and find that the cP
values decrease with time, the material is thixotropic.
If you take rheometer readings from a low RPM to a high RPM and then back to the
low RPM, and the readings are lower for the descending step, the material is time
dependent, thixotropic.
D V N E X T R H E O M E T E R - O P E R AT I O N A L M A N U A L
95