Storing the settings permanently with Bootp

Storing the settings permanently with Bootp

If you want to change parameters of the module using the Bootp follow the
instructions below.
1 Turn off the module.
2 Change the module's settings of the Configuration Switch to "Bootp &
3 Start the Agilent Bootp Service and open its window.
4 If required, modify the parameters for the module according to your needs
5 Press OK to exit the Bootp Manager.
6 Now turn on the module and view the Bootp Server window. After some
7 Close the Agilent Bootp Service and turn off the module.
8 Change the settings of the module's Configuration Switch to "Using Stored"
9 Power cycle the module. The module can be accessed now via LAN without
76
Store" mode, see
Table 16
using the existing configuration.
time the Agilent Bootp Service will display the request from the LAN
interface. The parameters are now stored permanently in the non-volatile
memory of the module.
mode, see
Table 16
on page 61.
the Agilent Bootp Service.
on page 61.
Agilent 1260 Infinity DAD and MWD User Manual