Event Messages
Display Text
W7003-7:EDP Error
W7003-8:EDP Error
W7003-nnn:EDP Error
**)
M7501:DISONET faulty
M7501-N:DISONET faulty
M7513-nnn:DISONET faulty**)
A7801:DISONET faulty
A7813-nnn:DISONET faulty**)
A8916-FFLLLLLEE:Run-time error
A8917-FFLLLLLEE:Run-time error
A9013-nnn:Parameter error
A9213-nnn:Parameter error
M9113-nnn:No data comparison at the time being
S9309:Battery voltage, defaults activated
A9411-nnn:Parameter error
A9511-nnn:Deadload Error
A9512-nnn:Deadload Error
A9611-nnn:Parameter Error
A9612-nnn:Parameter Error
A9613-nnn:Parameter Error
A9614-4:Parameter Error
A9711-nnn:Parameter Error
A9712-nnn:Parameter Error
A9713-nnn:Parameter Error
A9714-1:Parameter Error
A9714-4:Parameter Error
S9811:Parameter Error
®
DISOMAT
Opus – Operating Manual
© Schenck Process GmbH, Darmstadt
Edit *) Cause
*
*
*
*
*
**)
**)
**)
**)
**)
**)
**)
**)
**)
**)
**)
**)
List of Event Messages
Cyclic message transmission
timeout
Cyclic transmission, no timer
Internal error in sending
response message
DISONET Slave: connection to
Master
DISONET Master: connection
to Slave N
MULTIRAIL transmission error:
Slave twin-unit weight
DISOBOX: DISONET Slave:
connection to Master
MULTIRAIL transmission error:
zero value to twin-unit slave
in SMALL foreground process:
file FF, line LLLLL, error EE
in SMALL background process:
file FF, line LLLLL, error EE
The twin-unit calibration
parameters stored in dongle
are different
MULTIRAIL: big twin-unit scale
calibration parameters
Twin-unit calibration
parameters in dongles cannot
be compared at the time being.
Figures not read out yet
Ex-variant: battery voltage too
low after longer cut-off.
Parameter default values
activated.
Data cannot be stored in
dongle
Acquire Scale 1 deadload
Acquire Scale 2 deadload
Read Scale 1 calibration
parameters
Read Scale 2 calibration
parameters
Read Twin-unit calibration
parameters
Read data record in SMALL
Write Scale 1 calibration
parameters
Write Scale 2 calibration
parameters
Read Twin-unit calibration
parameters
Write data record in SMALL
Write data record in SMALL
Measuring module 1 improperly
calibrated
0630
Remedy
Edit calibration
parameters or
plug appropriate
dongles.
Edit calibration
parameters or
plug appropriate
dongles.
Edit calibration
parameters or
plug appropriate
s
dongle
Edit calibration
parameters or
plug appropriate
dongles.
Repeat at a later
time
Re-acquire
parameters.
Exchange dongle
Check dead load
or exchange
converter card.
Check dead load
or exchange
converter card.
Exchange
measuring
module.
BV-H2313 GB