Utility Requirements; Electrical Requirements; Instrument Air Supply; Sample Temperature And Pressure Requirements - Ametek 5800 User Manual

Moisture
Table of Contents

Advertisement

3-4 | 5830 Moisture Analyzer

Utility Requirements

Electrical requirements

The 5830 Moisture Analyzer must be installed in a general purpose area. It
does not have an On/Off power switch so power must be applied through
a circuit breaker switch at the AC source.
General Purpose installation:
Power connection can be made using the supplied power cord(s) ter-

minated with a locally approved wall socket plug, or
the analyzer can be hardwired.


Instrument air supply

Use dry instrument air ( 70 to 100 PSI) that is free of oils and other
contaminants for valve operation. N
-
If there is any doubt about the cleanliness of the air supply, con-
nect it through oil / air separators.
Air tubing must be fitted to a 1 /4 inch Swagelok® fitting.

Sample Temperature and Pressure Requirements

Inlet sample temperature:
Optimum input gas temperature:
For optimum performance, sample line should be heat traced to main-
tain a constant sample temperature.
NOTE
High Pressure :
Minimum input pressure: 138 kPa (20 PSIG)
Maximum pressure: 345 kPa (50 PSIG)
Low Pressure: (See specific gas table below)
Minimum input pressure: 34.5 kPa (5 PSIG)
Maximum pressure: 138 kPa (20 PSIG)
can be used as an alternate.
2
0 °C to 100 °C ( 32 °F to 212 °F )
60 °C ± 5 °C ( 140 °F ± 9 °F )