Sensitivity; Maximum Water Content; Customer Connections; Analog Outputs - Ametek 241CE II Operator's Manual

Hydrocarbon dewpoint analyzer
Hide thumbs Also See for 241CE II:
Table of Contents

Advertisement

Sensitivity

Maximum Water Content

Customer Connections

Electrical Requirements

2-2 | Model 241CE II Hydrocarbon Dewpoint Analyzer
± 0.1 °C (± 0.2 °F)
This is for the stability of temperature measurement only.
3
500 mg/m
(equivalent to a +5 °C water dewpoint at 200 PSIG)

Analog Outputs

4 isolated 4–20 mA
DC
Self-Powered:
Configured as a 2-wire transmitter, compatible with a loop impedance of
0–1000 ohms.
Loop-Powered:
1500 Vrms isolated analog output configured as a 2-wire transmitter. An
external power supply (15–30 VDC) capable of supplying 25 mA per loop
is required (supplied by customer). When 30 VDC is supplied, the maxi-
mum loop impedance is 1000 ohms.
6 independent sets of SPDT Form C, dry (potential free) relay contacts,
24 VDC, 0.5 A Maximum

Digital Communication

Local Service Port:
Remote Service Port: RS-422/RS-485 (2-wire or 4-wire) using the optional

Power Consumption

< 275 W
Maximum Current:

Supply Voltage

120 VAC, 50/60 Hz, ±10 % max.
240 VAC, 50/60 Hz, ±10 % max.
self-powered (standard) or loop powered (optional).
RS-232
serial communications, Modicon Modbus
col, or using the optional System 200 Configurator
Software installed on a laptop computer.
3 A
®
proto-