Condition Errors 501 to 599, Frequency

This series of errors corresponds to the bits in the
STATus:QUEStionable:FREQuencyr register. The second column in the table
shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 503 indicates a Frequency Reference Unlocked condition
has been detected, error 1503 indicates that failure has been cleared.
This register is summarized as bit 5 of the STATus:QUEStionable register. See
section
Err#
Bit in
status
register
501
0
503
1
505
2
507
3
509
4
511
5
513
6
515
7
517
8
519
9
521
10
523
11
525
12
527
13
529
14
142
"Condition Errors 601 to 699, Error Summaries" on page
Message
Source Synth Unlocked
Frequency Reference
Unlocked
nd
2
LO Unlocked
unused
LO Unlocked
unused
IF Synthesizer Unlocked
Calibration Oscillator
Unlocked
unused
Demodulation
External ref missing or
out of range
unused
unused
unused
unused
Error or
More Information
Warning
E
E
E
E
E
E
E
E
The external frequency reference signal
is missing or is not within the proper
amplitude range.
In the PSA, this error was reported as
error 622
N9030A PXA Signal Analyzer Service Guide
143.

Advertisement

Table of Contents
loading

Table of Contents