Assembly Replacement Procedures
RF Area (Options 543, 544, 550)
Replacement
1. Refer to
2. Place the switch/bracket into place into the chassis and replace the two
3. Refer to
4. Replace the ribbon cable W7. Ensure locking tabs on the sides of the
5. Refer to
6. Replace the front panel. Refer to the
7. Replace the instrument outer case. Refer to the
N9030A PXA Signal Analyzer Service Guide
Figure
16-47. Place the switch onto the bracket and replace the
five screws (3) (0515-0372). Torque to 9 inch-pounds.
screws (0515-0372). Torque to 9 inch-pounds starting with the screw
closest to the front of the instrument.
Figure
16-45. Replace the semi-rigid cables W6, W8, and W56.
Torque to 10 inch-pounds.
connector are engaged.
Figure
16-43. Position the RF bracket onto the chassis and
replace the sixteen screws (0515-0372). Torque to 9 inch-pounds.
procedure.
replacement procedure.
Front Frame Assembly
Instrument Outer Case
replacement
467