Keysight Technologies Option 503 Service Manual page 543

Pxa signal analyzer
Table of Contents

Advertisement

Post-Repair Procedures
Post-Repair Procedures
Table 17-1
Post-Repair Testing Requirements
Ad justments to Perform
Assembly
(in the order listed)
A1A2
None
Front Panel
Interface
A1A3
None
Liquid Crystal
Display
A2
IF Flatness Adjust (Option B25, B40, or B1X if
Analog IF Assembly
Option MPB is also present)
A3
FPGA Synchronization
Digital IF Assembly
A4
BIOS Settings Verification
CPU Assembly
Set Instrument Time and Date
A4BT1
BIOS Settings Verification
Battery
Set Instrument Time and Date
A5
Accept End User License Agreement (EULA)
d
Disk Drive
Instrument Software Update
All Adjustments
A6
None
Power Supply
A7
Program Instrument Model and Serial
Midplane board
Number
A8
None
Motherboard
N9030A PXA Signal Analyzer Service Guide
a
b
b
c
b
c
e
d
f
b
Performance Verifications to Perform After
a
Ad justments
Residual Responses
Displayed Average Noise Level
Residual Responses
Displayed Average Noise Level
Power Bandwidth Accuracy
Resolution Bandwidth Switching Uncertainty
Residual Responses
Displayed Average Noise Level
IF Frequency Response (Option B25, B40, B1X)
Absolute Amplitude Accuracy
Display Scale Fidelity
Power Bandwidth Accuracy
Resolution Bandwidth Switching Uncertainty
Residual Responses
Displayed Average Noise Level
IF Frequency Response (Option B25, B40, B1X)
Absolute Amplitude Accuracy
Display Scale Fidelity
Residual Responses
Displayed Average Noise Level
Residual Responses
Displayed Average Noise Level
Residual Responses
b
Displayed Average Noise Level
Frequency Response (All Tests)
Residual Responses
Displayed Average Noise Level
Spurious Responses
Residual Responses
Displayed Average Noise Level
Run all tests since instrument was disassembled.
543

Advertisement

Table of Contents
loading

Table of Contents