Keysight Technologies Option 503 Service Manual page 113

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–222
Data out of range; Two
entries already exist at
this x-axis value.
–223
Too much data
–223
Too much data; 200 spurs
found. Additional spurs
ignored.
–224
Illegal parameter value
–224
Illegal parameter value;
<Value> invalid.
Fractional values are not
allowed.
–224
Illegal parameter value;
<value> out of range.
–224
Illegal parameter value;
Exceeding the max list
length
–224
Illegal parameter value;
Gated FFT is not available
while Sweep Type is set to
Swept
–224
Illegal parameter value;
Gated LO is not available
while Sweep Type is set to
FFT
–224
Illegal parameter value;
Gated Video is not
available while Sweep Type
is set to FFT
–224
Illegal parameter value;
Illegal identifier
<identifier>. This value
may already be in use.
–224
Illegal parameter value;
Index out of range
N9030A PXA Signal Analyzer Service Guide
Verbose/Correction Information
When entering values for limit lines, you cannot have more than
two y-axis (amplitude) values entered for a specific x-axis
(frequency) value.
A data element (of block, expression, array type, or string type)
had more data then allowed by the command, or by the available
memory.
There are too many spurs for the table (the limit is 200), and any
additional spurs that are found will be ignored.
An exact data value (from a list of the allowed values) was required
- but not found. See the feature description for information about
the expected parameter values.
The seconds parameter of an LXI time may not contain a fractional
portion. For example 123456789.0 is valid while 123456789.1 is
not.
The value does not fall in the valid range
The list parameters have a maximum allowed length. You are
trying to set a length longer than the maximum.
The gated FFT function is not available if you have selected the
swept type of sweep. You must be in the FFT sweep type.
The FFT sweep type moves the LO frequency in steps. So the
gated LO function is not available if you have selected FFT sweep.
The gated video function is not available if you have selected the
FFT sweep type.
The value for the LXI LAN identifier parameter must be unique (i.e.
LAN0 and LAN7 must have different identifier strings).
When querying the LXI Event Log or the Servo Log, an index may
be used to look at a specific entry. This error occurs if the index
provided does not point to a valid entry.
113

Advertisement

Table of Contents
loading

Table of Contents