Keysight Technologies Option 503 Service Manual page 504

Pxa signal analyzer
Table of Contents

Advertisement

Assembly Replacement Procedures
Disk Drive
5. Remove the disk drive (2) from the carrier assembly.
6. Place the new disk drive onto the carrier assembly and attach with the 4
7. Refer to
8. Locate the Windows License Sticker from the kit and apply the sticker to
Figure 16-82
Windows License Sticker Rear Panel Location
9. Refer to
504
machine screws (1) (0515-0372). Torque to 9 in-pounds.
Figure
16-80. Replace the disk drive carrier assembly (2) into the
instrument and tighten the screw to 9 in-pounds.
the rear of the instrument as shown in
license sticker.
Chapter 17, "Post-Repair Procedures."
specified tasks for replacing the A5 Disk Drive assembly.
Figure
16-82, covering the existing
and perform all of the
N9030A PXA Signal Analyzer Service Guide

Advertisement

Table of Contents
loading