Functional Tests

Frequency Response (Flatness)

Frequency Response (Flatness)
Test Limits
Frequency Range
20 Hz to 3.6 GHz
> 3.6 GHz to 8.4 GHz
> 8.4 GHz to 13.6 GHz
> 13.6 GHz to 26.5 GHz
> 26.5 GHz to 34.5 GHz
> 34.5 GHz to 50 GHz
The frequency response test measures the signal analyzer's amplitude error as
a function of the tuned frequency. Measurements are made ranging from 50
MHz to the maximum frequency range of your analyzer. The signal source
amplitude is measured with a power meter to eliminate error due to source
flatness. The measured value is normalized to 50 MHz.
Item
Adapter
Type N (m) to 3.5 mm (m)
Adapter,
3.5 mm (f) to 3.5 mm (f)
Cable, 3.5 mm, 1 meter
(Opt 503, 508, 513, 526)
Cable, 2.4 mm, 1 meter
(Opt 543, 544, 550)
Cables, BNC
120 cm
Power Meter
Power Sensor
(Opt 503, 508, 513, 526)
Power Sensor
(Opt 543, 544, 550)
N9030A PXA Signal Analyzer Service Guide
Limit Relative to 50 MHz
±1.5 dB
±2.5 dB
±3.0 dB
±3.5 dB
±3.5 dB
±4.2 dB
Critical Specifications
(for this test)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Frequency: 10 MHz to 50 GHz
VSWR: < 1.55:1
Frequency: 10 MHz
Compatible with power sensor
Frequency Range: 50 MHz to 26.5 GHz
Frequency Range: 50 MHz to 50 GHz
Recommended
Keysight Model
1250-1743
83059B
11500E
8120-6164
10503A
E4418B
8485A
8487A