Condition Errors 301 To 399, Uncalibrated Integrity - Keysight Technologies Option 503 Service Manual

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Condition Messages

Condition Errors 301 to 399, Uncalibrated Integrity

This series of errors corresponds to the bits in the
STATus:QUEStionable:INTegrity:UNCalibrated sub-register. The second
column in the table shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 301 indicates a Meas Uncal condition has been detected,
error 1301 indicates that failure has been cleared.
This register is summarized as bit 3 of the STATus:QUEStionable:INTegrity
register. See section
page
Err#
Status
Register
Bit
301
0
303
1
305
2
307
3
309
4
N9030A PXA Signal Analyzer Service Guide
"Condition Errors 101 to 199, Measurement Integrity" on
128.
Message
Meas Uncal
Signal ID on
No Long Code Phase
AC coupled: Accy unspec'd
<10 MHz
User cal
Error or
Verbose/Correction Information
Warning
W
W
W
W
AC input coupling will function at lower
frequencies, but the performance is not
specified below 10 MHz.
W