Keysight Technologies Option 503 Service Manual page 109

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–221
Settings conflict;Function
not available while
measurement is running
–221
Settings conflict;function
unavailable with this EMC
Standard
–221
Settings conflict;Log
Percent Auto Step Rule does
not apply in Swept Scan
Type.
–221
Settings conflict;QPD + EMI
Average + RMS Average is not
allowed
–221
Settings conflict;Range <0>
is turned off as total range
points > 40001
–221
Settings conflict;RF Input 2
is not available above 1GHz
–221
Settings conflict;Scan Time
& Points do not apply in
current Step/Time Control.
–221
Settings conflict;Step Size
& Meas Time do not apply in
current Step/Time Control.
N9030A PXA Signal Analyzer Service Guide
Verbose/Correction Information
Settings change is not allowed while measurement is
running. You must stop the current measurement if you wish
to change the settings.
The Log Percent rule is not available when Scan Type=Swept
Scan because we are always doing linear sweep.
User is not allowed to turn on all 3 EMI detectors together.
You must turn off one of the EMI Detectors before you turn
this on.
Max of Total range points is 400000. Reduce Scan Points or
increase Step Size in order to turn on that range.
If the frequency range is set above 1GHz, user is not allowed
to change to RF Input 2.
The Scan Time & Points are not available when Step/Time
Control is set to Step & Dwell.
The Step Size & Meas Time are not available when
Step/Time Control is set to Scan Time&Pts.
109