Instrument Messages
Event Messages
Err#
Message
–221
Settings conflict; Gate is
not available when Marker
Count on
–221
Settings conflict; Gate
Length is not settable in
FFT sweeps
–221
Settings conflict; Gate
Length is not settable in
FFT sweeps
–221
Settings conflict; Gate
Method is not compatible
with current Sweep Type
setting
–221
Settings conflict; Gate not
available with external
Tracking Source
–221
Settings conflict; Gate not
available with Tracking
Generator
–221
Settings conflict; Gated FFT
is not available while Sweep
Type is set to Swept
–221
Settings conflict; Gated LO
is not available while Sweep
Type is set to FFT
–221
Settings conflict; Gated
Video is not available while
Sweep Type is set to FFT
–221
Settings conflict; Incorrect
RBW for demod. Change RBW
–221
Settings conflict; Ind I/Q
is not available for this
measurement
–221
Settings conflict; Invalid
trace number
–221
Settings conflict; Knob is
not available to modify this
function
–221
Settings conflict; LO Phase
Noise Adj not available
N9030A PXA Signal Analyzer Service Guide
Verbose/Correction Information
The gate function cannot be used while you have marker
count turned on.
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
If the Gate is On and you have the FFT Sweep Type manually
selected, then the Gate Method cannot be selected.
The Gate functions are unavailable when Source Mode is
Tracking with an external source. This is because the Gate
circuitry is used to sync the external source.
If the Source Type is Tracking Generator, the Gate circuitry is
used for TG sync and is not available for gating
The gated FFT function is not available if you have selected
the swept type of sweep. You must be in the FFT sweep type.
The FFT sweep type moves the LO frequency in steps. So the
gated LO function is not available if you have selected FFT
sweep.
The gated video function is not available if you have selected
the FFT sweep type.
The Independent I and Q setting is not available for the
current measurement. Only some measurements (initially,
only VXA) support this setting.
The subopcode used to specify the trace number is invalid
for this measurement or query
You should select a specific value for this function. So
scrolling through values with the knob is not allowed.
For instruments without the Dual-Loop LO, this feature is not
available
103