Instrument Messages
Event Messages
Err#
Message
–200
Execution Error;
Preselector centering
failed
–200
Execution Error; Signal
not stable enough to track
–200
Execution Error; Store ref
trace before turning on
Normalize
–200
Execution error; Sync word
was not found.
–200
Execution error; Trace
file contains no
compatible traces.
–200
Execution error; Trace
file created by
incompatible version of
Phase Noise App
–200
Function not available
before Marker Zoom is
performed
–200
Must perform Scan before
do Search
–200
No marked signal
–200
No Measure At Marker Added
to Signal List
–200
No Peak Added to Signal
List
–200
Signal List is Empty
–200
Signal List is Full
–200
Signal Selected is not in
the Signal List
N9030A PXA Signal Analyzer Service Guide
Verbose/Correction Information
Algorithm failed to center the preselector. This maybe caused by
the signal peak being too low in amplitude. Or it could be from
excessive CW input signal, alignment error, or hardware failure.
The signal that you have selected to track is changing too much
for the function to track it properly.
The Reference trace data must be stored in the Ref trace before
you turn on the Normalization function.
NADC & PDC: In an EVM measurement, the sync word is not
found and the synchronization cannot be established when Sync
Word is selected in the Burst Sync menu.
Flexible Digital Demodulation: The sync word cannot be detected
because of inappropriate parameter settings or incorrect signal.
The trace file may have been created by another version of the
Phase Noise personality, which uses a different trace format that
is incompatible with the version you are running. Please check
you are running the most up to date version of the personality.
The trace file may have been created by another version of the
Phase Noise personality, which uses a different trace format that
is incompatible with the version you are running. Please check
you are running the most up to date version of the personality.
Function not available before perform Marker Zoom
Cannot perform Search as the trace data is found empty
Cannot perform the selected function because no signal was
marked. You must mark the peak of interest before selecting the
function.
No signal peak was added into Signal List as there is no valid
measure to Marker result. You must perform Measure at maker
before selecting Measure At Marker --> List.
No signal peak was added from the Search to Signal List as there
is no signal peaks found within the defined parameters of the
search criteria.
Cannot perform the selected function because the signal list
contains no data.
Cannot perform the selected function because the signal list is
full. Please clear the list.
The signal selected is not the list. You only can perform the
operation on signal that is already available in the list.
111