  • Page 1 Keysight Technologies N9030B PXA Signal Analyzer Option B85, 25 MHz to 85 MHz Analysis Bandwidth Upgrade Option B1X, 25 MHz to 160 MHz Analysis Bandwidth Upgrade Option BU3, 40 MHz to 160 MHz Analysis Bandwidth Upgrade Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Refer to http://www.keysight.com/find/pxa_upgrades for information on all available upgrades In addition to installing the licenses and hardware to support options B85, or B1X, as appropriate, the Option B85 and Option B1X kits also include a license for N9030B-B40. Installation Note N9030-90075...
  • Page 4 Option B85 or B1X Retrofit Kit The B85 or B1X option is licensed for only one instrument model number/ serial number combination. The license file that is downloaded from the web will only install on the designated instrument. The instrument must be readjusted and the performance tested to assure the instrument meets specifications following the hardware installation.
  • Page 5: Installation Kit Parts List

    Option B85 or B1X Retrofit Kit Installation Kit Parts List Quantity Description Keysight Part Number Installation Note This note Option Upgrade Entitlement Certificate 5964-5178 Wideband Analog IF Assembly N9020-60044 Wideband Digital IF Assembly N9020-60311 Cable Assembly, Flat Flexible 80-Conductor 3-in-LG 8121-1854 Screw, M3 x 0.5, 6 mm, Flathead 0515-1946...
  • Page 6: Tools Required

    Option B85 or B1X Retrofit Kit Tools Required — T-10 TORX Driver — T-20 TORX Driver — 5/16-inch torque wrench — Keysight Calibration and Adjustment Software, N7814A TME Calibration Application, version E.16.00 or later — Test equipment and computer supported by the X- Series Performance Tests and Adjustment Software —...
  • Page 7 Option B85 or B1X Retrofit Kit Procedure If the instrument is placed on its face during any of the following procedures, be sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or input connector. Make sure any adapters on the front panel are removed.
  • Page 8 2. Double-tap on the LaunchXSA icon on the Windows desktop. Wait for the XSA application to finish starting (the analyzer should be sweeping). 3. Press System, Show System — N9030B-B85 Analysis bandwidth, 85 MHz (if N9030BU-B85 is being installed) — N9030B-B1X...
  • Page 9 Option B85 or B1X Retrofit Kit Analyzer Disassembly 1. Refer to the Assembly Replacement Procedures chapter in the N9030B Signal Analyzer Service Guide 2. Perform the removal procedures shown below in the order listed a. Instrument Outer Case b. Top Brace (discard flathead screws with patch lock) c.
  • Page 10 Option B85 or B1X Retrofit Kit Removing Cables 1. Refer to Figure 5. Remove wire cable hold down on right side of chassis to free gray coax cables. 2. Remove cable ties from bundle of gray coax cables. 3. Refer to Figure 3 that shows the A3 Digital IF assembly located on the bottom side of the instrument.
  • Page 11 Option B85 or B1X Retrofit Kit Installing Boards and Cables 1. Refer to Figure 4. Install the Wideband Analog IF assembly into slot 3. 2. Install the Wideband Digital IF into slot 5. Slot 4 must remain empty. Figure 4 Board Installation Installation Note N9030-90075...
  • Page 12 Option B85 or B1X Retrofit Kit 3. Connect the flat ribbon cable 8121-1854 between the Wideband Analog IF assembly and the Wideband Digital IF assembly. Figure 5 Option Cables 4. Select the coax cable with color bands 102 and 15 from the kit. Connect the cable end with color code 102 to Wideband Analog IF assembly J102.
  • Page 13 Option B85 or B1X Retrofit Kit Analyzer Reassembly 1. Refer to the Assembly Replacement Procedures chapter in the N9030B Signal Analyzer Service Guide. 2. Perform the replacement procedures shown below in the order listed: a. Right Side Chassis b. Top Brace (use the new flathead screws supplied with this kit) c.
  • Page 14 Option B85 or B1X Retrofit Kit Power Up and New Hardware Wizard 1. Connect a keyboard and mouse to the instrument. 2. Power on the instrument. 3. During the boot up process you may notice that the “Found New Hardware” bubble appears in the lower right screen, and a “Found New Hardware’...
  • Page 15 Option B85 or B1X Retrofit Kit 6. The screen in Figure 7 appears. Select Yes, This time only. Click Next. Figure 7 Installation Note N9030-90075...
  • Page 16 Option B85 or B1X Retrofit Kit 7. The screen in Figure 8 appears. Ensure “Install the software automatically’ is selected and click Next. Figure 8 8. The wizard will install the required software. Once you see the “Completing the Found New Hardware Wizard”...
  • Page 17: Verify Hardware Installation

    Option B85 or B1X Retrofit Kit Verify Hardware Installation 1. Verify the spectrum analyzer application loads and sweeps as expected. 2. Press System, Show Hard ware on the analyzer and verify that the new board assemblies identify themselves as follows: —...
  • Page 18 Perform all performance tests End of installation. For assistance, get in touch with your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...
  • Page 19 This information is subject to change without notice. © Keysight Technologies 2016 Edition 1, March 2016 N9030-90075 www.keysight.com...