Instrument Messages
Event Messages
Err#
Message
–310
System error; No
license; <feature
code> will terminate
in <time>
–310
System Error; No
supported source
found
–310
System Error; source
connection lost,
check interface
connection
–310
System error; The
configured PTP
hardware driver could
not be instantiated.
–310
System error; The PTP
hardware driver
reported a
configuration error.
–310
System error; The PTP
ordinary clock
reported a
configuration error.
–310
System error; The
Trigger alarm delayed
LAN event was not
scheduled due to an
existing pending
event.
–310
System error; The
Trigger alarm delayed
LAN event was not
scheduled due to a
conflict with an
existing scheduled
alarm.
–310
System error; The
Trigger alarm was not
scheduled due to a
conflict with an
existing scheduled
alarm.
–311
Memory error
N9030A PXA Signal Analyzer Service Guide
Verbose/Correction Information
The specified feature will stop working in the specified time due to the
license expiration You will be prompted to save results and exit.
Signal source at given IP address is not responding / IP does not belong to
a source. Check IP address and network connection.
Signal source at given IP address is not responding / IP does not belong to
a source. Check IP address and network connection.
The PTP driver failed on initialization.
Failure in the execution of the PTP driver. The most likely cause of this error
is a mismatch between versions of the PTP driver and the LXI middleware.
Failure in execution of the DMC libraries' PTP controller.
Delayed LAN events cannot occur too close together (within 20 ms).
Delayed LAN events cannot occur too close to a scheduled Alarm (within
20 ms).
Alarms cannot be scheduled to happen too close together (within 20 ms).
There is a physical problem with the instrument memory, such as a parity
error.