Instrument Messages
Condition Messages
Thus, the summary bits cannot be used to determine the current state of a
lower level condition bit; only the state and history of the lower level event bits.
This register is itself summarized as bit 8 of the STATus:QUEStionable register.
See section
Err#
Bit in
status
register
36
0
38
1
40
2
42
3
44
4
46
5
48
6
50
7
52
8
54
9
56
10
58
11
60
12
62
13
64
14
N9030A PXA Signal Analyzer Service Guide
"Condition Errors 601 to 699, Error Summaries" on page
Message
unused
unused
TG Alignment Failure
RF Alignment Failure
IF Alignment Failure
LO Alignment Failure
ADC Alignment Failure
FM Demod Alignment
Failure
Extended Align Needed
Summary
Extended Align Failure
Summary
unused
Align Skipped Sum Summary
Align Now, RF required
unused
Align Now, All required
Error or
More Information
Warning
E
E
E
E
E
E
status bit
This bit is the summary bit for the
only
STATus:QUEStionable:CALibration:EXTende
d:NEEDed sub-register.
status bit
This bit is the summary bit for the
only
STATus:QUEStionable:CALibration:EXTende
d:FAILure sub-register.
status bit
This bit is the summary bit for the
only
STATus:QUEStionable:CALibration:SKIPped
sub-register.
E
E
On PSA, this was error 64