Service and Diagnostics Menus
Service Key Descriptions

Align

The Align Menu allows accessing Diagnostic capabilities of Alignment, and
invoking alignments for individual subsystems.
358
Diagnostic
The Diagnostic menu contains items for controlling the operating behavior
of Alignment and the Alignment reporting capabilities.
Visible Align
Controls the state of Visible Align. Visible Align replaces the "Aligning
1 or X" messages with descriptive messages describing the
alignment piece begin executed.
Align Log...
Invokes Notepad with the Alignment Log loaded. The log can be
viewed or saved to an external media or drive. Notepad can be closed
without a mouse or external keyboard by pressing ALT front-panel
key, then arrow down to highlight Exit, then press Enter.
Align Log Mode
The Alignment System places information about the last alignment
performed into a log. The log can be configured to operate in clear or
append mode. In clear mode the log only contains information on the
last alignment performed; the log is cleared with each alignment
performed. In append mode each new alignment appends to the log.
In append mode, care must be taken by the operator to not allow the
log to grow to a size so large as to inhibit system operation.
Subsystem
The Subsystem menu allows invoking individual components of the internal
alignments.
ADC
Immediately executes an alignment of the ADC subsystem. The
instrument will stop any measurement currently underway, perform
the alignment, then restart the measurement from the beginning.
A failure of ADC will set the Error Condition "Align ADC failed". A
failure will not employ new ADC alignment data.
Successful completion of ADC will clear the Error Condition "Align
ADC failed".
The Advisory Event "Alignment complete" is displayed when the
alignment is complete.
Align ADC can be interrupted by pressing the Cancel (ESC)
front-panel key. When this occurs, no new ADC alignment data will
be employed.
N9030A PXA Signal Analyzer Service Guide