Instrument Messages
Event Messages
Err#
Message
–221
Settings conflict; Trace
Math is not available while
Normalize is on
–221
Settings conflict; Tracking
Source unavailable in FFT
Sweeps
–221
Settings conflict; Trigger
input in use for source
synchronization
–221
Settings conflict; Trigger
is not available with span >
0 Hz.
–221
Settings conflict; Tx Band
Spur meas does not support
this frequency band.
–221
Settings conflict; Tx Band
Spur measurement is not
defined for mobiles.
–221
Settings conflict; Zero Span
not available when Frequency
Scale Type = Log
–221
Settings
conflict;Administrator
privileges required
–221
Settings conflict;Auto Scan
Time/Meas Time do not apply
in Stepped Scan Type.
–221
Settings conflict;Auto Scan
Time/Meas Time do not apply
in Stepped Scan Type.
–221
Settings conflict;ClearList
& Start not available when
ScanSeq = (Re)measure
–221
Settings conflict;EMI
Detectors and Average
detector can't be used
together
–221
Settings conflict;Freq > 1
GHz is not available while
RF Input 2 enabled
108
Verbose/Correction Information
The Normalize function works by doing trace manipulation,
so trace math is not available while normalization is running.
Since FFT's do not sweep, you cannot use a Tracking Source
while doing FFT's
If Point Trigger is being used with an external trigger input to
synchronize an external source to the analyzer, that trigger
input is unavailable for triggering.
The transmit band spur measurement does not support all of
the commercially available frequency bands. You need to
change your selection under Mode Setup, Radio, Band to
one of the supported bands.
Only base station testing is available.
Logarithmic scales cannot be used for time domain sweeps
(0 Hz span).
You must be logged in with administrator privileges to do
this. Log out and log back in as the administrator, then
restart the SA application.
The Auto Scan Time/Meas Time are not available when Scan
Type =Stepped Scan.
The Auto Scan Time/Meas Time are not available when Scan
Type =Stepped Scan.
Cannot perform Clear List & Start during (Re)measure
because we needs the frequency information of the peaks in
signal list to perform (Re)measure.
User is not allowed to turn on any EMI Detector and Average
detector together. They are always mutually exclusive.
Frequency is limited to 1GHz while RF Input 2 is enabled.
N9030A PXA Signal Analyzer Service Guide