Assembly Replacement Procedures
Front Frame Assembly
Inverter Board Assembly
Removal
1. Refer to
2. Disconnect the wires (1) from the Inverter board (3).
3. Remove the two screws (2) (0515-0372) securing the Inverter board (3).
Figure 16-101
Inverter Board Removal
Replacement
1. Place the Inverter board into the inverter bottom shield and position onto
2. Reattach the wires to their correct locations.
3. Place the inverter top shield into position. Replace the eight screws (32)
N9030A PXA Signal Analyzer Service Guide
Figure
16-100. Remove the eight screws (32) (0515-0372) and
the inverter top shield.
The Inverter board and inverter bottom shield can now be removed.
the display bracket. Replace the two screws (0515-0372). Torque to
9 inch-pounds.
(0515-0372) as shown in
Figure
16-100. Torque to 9 inch-pounds.