IF Flatness
This function turns the corrections related to IF flatness On or Off.
When IF Flatness is OFF, the Advisory Event "IF Flatness corrections OFF"
will be displayed.
When IF Flatness is ON, the Advisory Event "IF Flatness corrections OFF"
will be cleared.
Snapshot Alignments On/Off
During normal auto alignment operation, the firmware will perform an RBW
and IQ alignment when the analyzer RBW state is changed, and every 10
minutes thereafter. These alignments can hinder troubleshooting the
analyzer since they can change correction factors when the state is
changed. Turning snapshot alignments off disables the RBW portion of
these alignments for more accurate troubleshooting. (Use IF Flatness
ON/OFF to control the IQ alignment.)
When Snapshot Alignments are OFF, the Advisory Event "Snapshot OFF"
will be displayed.
When Snapshot Alignments are ON, the Advisory Event "Snapshot OFF"
will be cleared.
I/Q Freq Response
This function turns the corrections related to I/Q Frequency Response On or
Off. This is only available with the Baseband Analog I/Q option (BBA).
When I/Q Frequency Response is OFF, the Advisory Event "I/Q Freq
Response corrections OFF" will be displayed.
When I/Q Frequency Response is ON, the Advisory Event "I/Q Freq
Response corrections OFF" will be cleared.
I/Q Gain
This function turns the corrections related to I/Q Gain On or Off. This is only
available with the Baseband Analog I/Q option (BBA).
When I/Q Gain is OFF, the Advisory Event "I/Q Gain corrections OFF" will be
displayed.
When I/Q Gain is ON, the Advisory Event "I/Q Gain corrections OFF" will be
cleared.