Boot Up and Initialization Troubleshooting
PXA Instrument Boot Up Process
4. The instrument fans should start running. The fans are mounted on the
5. The Keysight Technologies splash screen is displayed in white font on a
6. Verify text is displayed on screen where the user has the option of booting
7. If the recovery system is not selected the Windows XP Professional
If a recovery was selected follow the on-screen instructions and perform a
system recovery. Additional information about performing a system
recovery can be found in
page 277
8. The Keysight Technologies logo is displayed in white font on a blue
9. By default, the initialization process of the PXA Spectrum Analyzer
10.While the application software is loading the instrument will perform a
N9030A PXA Signal Analyzer Service Guide
left-hand side of the analyzer and draw air into the instrument to cool the
internal circuitry. If a fan is not running refer to the
Operating"
section in this chapter.
dark background for ~5-10 seconds after the analyzer is turned on. If the
Keysight Technologies logo is not displayed refer to the
Splash Screen Displayed"
at the Keysight Technologies splash screen refer to the
at the Keysight Splash Screen"
Windows XP or running the Instrument Recovery System. The default
selection is to boot Windows XP. If a recovery is required, press the Down
Arrow key on the front panel of the analyzer within 5 seconds to highlight
"Keysight Recovery System" and press the Enter key on the analyzer,
otherwise Windows XP will begin to boot. If the Windows XP boot screen
is not displayed within a few seconds refer to the
Completely Load or Run the Operating System"
operating system will begin to boot up. This will take ~20-30 seconds.
in this manual.
background while Windows finishes loading user preferences. This can
take up to 4 minutes. If this does not occur refer to the
Completely Load or Run the Operating System"
application begins loading. The N9030A PXA Signal Analyzer screen
appears. This screen remains for slightly over 1 minute. If any of the
initializing processes do not complete, refer to the
Complete"
section in this chapter.
number of internal alignments before the analyzer begins to sweep. A
screen with a black background appears and the alignment progress is
shown in a yellow box. The number of alignments is dependent on which
hardware options are present. If an alignment fails, refer to the
Initial Alignment"
section in this chapter.
section in this chapter. If the instrument hangs
section in this chapter.
Chapter 9, "CPU/Disk Drive Troubleshooting", on
"Fan(s) Are Not
"No Keysight
"Instrument Hangs
"Instrument Cannot
section in this chapter.
"Instrument Cannot
section in this chapter.
"Initializations Did Not
"Fails an