Assembly Replacement Procedures
RF Area (Options 543, 544, 550)
Replacement
1. Refer to
2. Place the switch/bracket assembly into the frame and secure with the two
3. Refer to
Figure 16-61
W7 Routing
N9030A PXA Signal Analyzer Service Guide
Figure
16-60. Attach switches (2)and (4) to bracket with three
screws
0515-1934
for each switch. Note that the switch with the wire
pigtail (4) mounts below the switch without the pigtail (2). Torque to
6 inch-lbs.
screws 0515-1035. Torque to 9 inch-lbs.
Figure
16-61. Route W7 in between the two switches as shown
and insert W7 connector into Low Band Switch header. Ensure the locking
tabs are engaged on both sides of the connector.