Keysight Technologies Option 503 Service Manual page 92

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Advisory Messages
Message
Reading SNS data...
Recalled File <filename>
Refer to online help for
assistance with DSB
measurements
Requested timeslot number is
not present.
RF Start freq should be
greater than IF Fixed freq
RF Start freq should be
greater than IF start freq
RF Start freq should be
greater than LO fixed freq
RF Start freq should be
greater than LO Start freq
RF Stop freq should be
greater than IF Stop freq
Scale/Div only applies in Log
Y Scale
Signal Track is turned off
when Zero Span is selected
Signals deleted
92
Description/Correction Information
The Keysight Smart Noise Source has been connected and the application
is reading the device EEPROM data. Please wait until complete before
continuing.
A file recall (open/load) was successfully completed.
The Double Side Band measurement requires careful setup to obtain valid
results. Please refer to the manuals for help with this setup.
The selected timeslot is not on. (Timeslot is referenced to the trigger
point.)
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF Start freq should be greater than the IF Fixed
freq for a DSB (for DSB measurements the setup uses LSB values)
downconverter swept LO setup. Use the graph icon on the DUT setup form
to clarify the setup required.
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF start freq should be greater than the IF Start
freq's for an LSB downconverter fixed LO setup. Use the graph icon on the
DUT setup form to clarify the setup required.
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF start freq should be greater than the LO fixed
freq's for an USB downconverter fixed LO setup. Use the graph icon on the
DUT setup form to clarify the setup required.
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF start freq should be greater than the LO Start
freq's for an USB downconverter swept LO setup. Use the graph icon on
the DUT setup form to clarify the setup required.
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The RF Stop freq should be greater than the IF Stop
freq's for an USB or DSB (for DSB measurements the setup uses LSB
values) downconverter fixed LO setup. Use the graph icon on the DUT
setup form to clarify the setup required.
Setting the Scale/Division only makes sense when you are using a
logarithmic Y scale.
Signal Track is not available when you have selected Zero Span. So if Zero
Span is entered while in Signal Track is On, Signal Track is turned off.
The signals in the signal list were successfully deleted.
N9030A PXA Signal Analyzer Service Guide