Keysight Technologies Option 503 Service Manual page 39

Pxa signal analyzer
Table of Contents

Advertisement

Overview
Service Equipment You Will Need
Table 1-1
Instrument
Cables
3.5 mm (m) to 3.5 mm (m)
(2 required)
2.4 mm (f) to 2.4 mm (m)
(for Options > 543)
Type-N
(2 required)
BNC
(3 required)
Filters
50 MHz Low Pass
300 MHz Low Pass
(2 Required)
1.8 GHz Low Pass
(2 Required)
N9030A PXA Signal Analyzer Service Guide
Required Test Equipment
Critical Specifications
Frequency: DC to 26.5 GHz
Length: ≤ 92 cm (36 in)
Insertion Loss: ~2 dB
VSWR: DC to 18 GHz: ≤ 1.25:1
18 GHz to 26.5 GHz: ≤ 1.35:1
Frequency: DC to 50 GHz
Length: ≤ 24.9 mm (9.8 in)
Insertion Loss:
≤ 26 GHz, ~4 dB
≤ 40 GHz, ~5 dB
≤ 50 GHz, ~6 dB
VSWR:
≤ 26.5 GHz: ≤ 1.30:1
≤ 40 GHz: ≤ 1.40:1
≤ 50 GHz: ≤1.55:1
Frequency: 10 MHz to 8 GHz
Precision Type-N (m), both ends
62 cm (24 in.)
VSWR: ≤ 18 GHz: 1.4:1
Insertion Loss: 1.5 dB
Frequency: DC to 10 MHz
50 Ω Coax BNC (m), both ends
120 cm (48 in.)
Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Insertion Loss: ~1 dB
VSWR: ≤ 1.5:1
BNC (m) to BNC (f)
Cutoff Frequency: 300 MHz
Rejection at > 435 MHz: > 45 dB
VSWR: ≤ 1.5:1
BNC (m, f)
Cutoff frequency: 1.8 GHz
Rejection at 2.1 GHz to 6 GHz: > 45 dB
Insertion Loss at 1.8 GHz: < 1 dB
VSWR: ≤ 1.50:1, DC to 1.8 GHz
Connector: SMA (f) to SMA (f)
Recommended
Al ternative
Keysight Model
Keysight Model
a
Number
or Vendor
Number
Model Number
11500E
8120-4921
8120-6164
11500C
10503A
Telonic Berkeley
0955-0306
TLA 50-5AB2
Telonic Berkeley
0955-0455
TLP 300-4AB4
RLC L-1636
0955-0491
b
Use
a
A, P
A, P
A, P, T
A, P, T
P,T
P
P
39