Keysight Technologies Option 503 Service Manual page 106

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–221
Settings conflict; Param
only available when
Frequency Mode is Swept
–221
Settings conflict; Param
only available when valid
cal data exists
–221
Settings conflict; Preamp
gain is not available in
this Mode
–221
Settings conflict; Preamp
unavailable with electronic
attenuator on
–221
Settings conflict; Reference
marker must be in same
window
–221
Settings conflict; Relative
Trigger needs hardware
support for this meas
–221
Settings conflict; Scale
Type = Lin is not available
when Normalize is on
–221
Settings conflict; Settings
conflict; Pre-trigger is
insufficient for demod.
Decrease Trig Delay.
–221
Settings conflict; Signal
Track is not available when
Freq Scale=Log
–221
Settings conflict; Signal
Track is not available with
Continuous Peak
–221
Settings conflict; Signal
Track is only available in
Swept SA measurement
–221
Settings conflict; Signal
Track is turned off when
Zero Span is selected
–221
Settings Conflict; Span
limited to XXX
106
Verbose/Correction Information
SCPI only message. This parameter is only available when
the Freq mode is set to Swept. Change the Freq Mode to
Swept
SCPI only message. The 'Apply Calibration' parameter is only
available when the stored cal data matches the current
setup. Perform a fresh 'Calibrate Now' or change setup such
that current cal data is valid.
Preamp gain correction is not available in some Modes or
Measurements
The electronic attenuator is on. Internal preamp cannot be
used while you are using the electronic attenuator.
A delta marker and its reference must be in the same
window. This error occurs when you try to turn on a delta
marker who's reference is in a different window.
To do Relative Triggering in this measurement requires
optional hardware that is not present in this analyzer.
Only the Log amplitude scale is available in Normalize, since
the results are always presented as a dB ratio.
The signal tracking feature cannot be used when you have
selected a log scale for the frequency axis.
The signal tracking feature cannot be used while you are also
using the continuous peak function.
The signal track functionality can be used when making a
swept SA measurement. It is not available in the SA
measurement when you are using FFT sweeps.
Signal Track is not available when you have selected Zero
Span. So if Zero Span is entered while in Signal Track is On,
Signal Track is turned off.
N9030A PXA Signal Analyzer Service Guide