Assembly Replacement Procedures
Front Frame Assembly
Figure 16-103
LCD and Display Bracket Separation
Replacement
1. Refer to
2. Replace the two screws (2) (0515-0372) that secure the display bracket
3. Replace the Inverter board.
N9030A PXA Signal Analyzer Service Guide
Figure
16-102. Place the display bracket into position in the Front
Frame Assembly. Reconnect the flex circuit A1W1 (2) to the front panel
interface board (4).
to the Front Frame Assembly. Torque to 9 inch-pounds.