Rf Section Description - Keysight Technologies Option 503 Service Manual

Pxa signal analyzer
Table of Contents

Advertisement

RF Section Troubleshooting (Millimeter-Wave Analyzers)

RF Section Description

RF Section Description
This section covers only those optional frequency ranges listed below for the
N9030A Signal Analyzer.
— Option 543, 43 GHz Frequency Range
— Option 544, 44 GHz Frequency Range
— Option 550, 50 GHz Frequency Range
The RF input signal can be routed through four different front end signal paths.
These various paths are switched in and out based on where the signal
analyzer center frequency is tuned.
1. Low band RF path is used when the input frequency is ≤ 3600 MHz
2. High band RF path #1 is used when the input frequency is > 3.6 GHz
3. High band RF path #2 is used when the input frequency is > 17 GHz to
4. High band RF Path #3 is used when the input frequency is > 34.4 GHz to
The RF section is designed to convert RF input signals to an intermediate
frequency that is present at the RF output of the A13 Front End assembly. This
intermediate frequency (IF) depends on installed options and the IF path
selected. This chapter will assume the default operation in Spectrum Analyzer
mode. Therefore the IF frequency will be 322.5 MHz.
174
to 17.1 GHz
34.5 GHz
50 GHz
IF Path
10 MHz (standard)
25 MHz (Option B25)
40 MHz (Option B40)
140 MHz (Option B1X)
160 MHz (Option B1X & B1Y)
IF
Mode
322.5 MHz
Spectrum Analyzer or IQ Analyzer
322.5 MHz
IQ Analyzer
250 MHz
IQ Analyzer
300 MHz
IQ Analyzer
300 MHz
IQ Analyzer
N9030A PXA Signal Analyzer Service Guide