Keysight Technologies Option 503 Service Manual page 458

Pxa signal analyzer
Table of Contents

Advertisement

Assembly Replacement Procedures
RF Area (Options 503, 508, 513, 526)
14.Refer to
Figure 16-39
W10, W14, W54 Installation
15.Install rigid cable
16.Install rigid cable
458
Figure
16-39. Install rigid cable
switch port 3 to Low Band Switch Assembly as shown. Torque to
10 inch-lbs.
Item
Keysight Part Number
W10
E4410-20163
W14
E4410-20165
W54
N9020-20145
W14
YTF Assembly as shown. Torque to 10 inch-lbs.
W10
YTF Assembly as shown. Torque to 10 inch-lbs.
W54
(E4410-20165) (2) from bottom switch port 2 to
(E4410-20163) (3) from middle switch port 2 to
N9030A PXA Signal Analyzer Service Guide
(N9020-20145) (1) from top

Advertisement

Table of Contents
loading

Table of Contents