RF Section Troubleshooting (Millimeter-Wave Analyzers)
Troubleshooting
Turn off auto align by pressing System, Alignments, Auto Align, Off.
IMPORTANT
Disconnect cable W16 at A13A1J7, 322.5 MHz output on the Front End
Assembly. See
Figure 5-4
A13A1J7 Location (Options 543, 544, and 550)
190
Figure
5-4. Select Span, 0 Hz.
N9030A PXA Signal Analyzer Service Guide