Assembly Replacement Procedures
Instrument Outer Case
Figure 16-3
Instrument Outer Case Removal

Replacement

1. Disconnect the instrument from ac power.
2. Slide the instrument cover back onto the deck from the rear. The seam on
3. Replace the four rear feet to the rear of the instrument. Torque the rear
4. Replace the bottom feet by sliding into place until they snap in. Install the
5. Replace the handle straps on both sides of the instrument. Torque the
N9030A PXA Signal Analyzer Service Guide
the cover should be on the bottom. Be sure the cover seats into the gasket
groove in the Front Frame Assembly.
feet screws
(0515-1619
locks by pressing in flat.
handle strap screws to 21 inch-pounds.
and
3050-0893
washers) to 21 inch-pounds.
423