Download Print this page

2-Port Insertion Loss Measurements - Keysight Technologies FieldFox B Series User Manual

Advertisement

CAT (Cable and Antenna Test) Mode

2-Port Insertion Loss Measurements

14.Use Averaging to remove random noise from high-loss measurements.
The displayed trace shows the Cable Loss values in one direction through the
cable. A Return Loss measurement would show the loss for both down the
cable and back. Therefore, a Cable Loss measurement is the same as a Return
Loss measurement divided by 2.
The average Cable Loss across the specified frequency range is shown on the
screen below the graticules.
2-Port Insertion Loss Measurements
A 2-port Insertion Loss measurement is used to measure the loss through a
DUT (device under test) – or cable – over a specified frequency range. The
FieldFox signal source is transmitted out the RF OUT connector, through the
DUT, and into the RF IN connector. Both ends of the DUT must be connected
to the FieldFox, either directly or indirectly using the cable used in the
normalization cal.
'Insertion' loss simply means loss through a device, usually expressed in dB. It
is exactly the same measurement as "S21 Transmission" in NA Mode.
Exception: In CAT mode, if the TDR data are saved in S1P format, the
values represent the real part of the complex transform; the imaginary part
is set to zero. To obtain complex data, either as real/imaginary or
magnitude/phase pairs, use NA mode.
2-port Insertion Loss measurements are generally more accurate than 1-port
Cable Loss measurements.
How to make a 2-port Insertion Loss Measurement
— Press Mode then CAT.
— Then
— Press Freq/Dist and enter
— Press Sweep 3, then select a
— Press Cal 5, then perform a calibration. Learn more on
— Connect the DUT and view the insertion loss measurement results.
When measuring very long lengths of cable, it may be necessary to increase
the sweep time. Learn how on
Help:
62
Press BW 2 then Average.
More
then
Insertion Loss
measurement.
Calibration" on page
127.
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
(2-Port).
Start
and
Stop
frequency values of the
Resolution
setting.
page
56. Learn why in the Supplemental Online
Keysight N9938-90006 User's Guide
"How to Perform a

Advertisement

loading