VVM (Vector Voltmeter) Mode - Option 308
VVM Mode Settings
This setting survives a Preset and Mode Preset.
How to set Resolution
— Press Scale/Amptd
— Then
— Then choose:
IF Bandwidth
Reducing the IF receiver bandwidth reduces the effect of random noise on a
measurement. However, narrower IF bandwidths cause longer measurement
times.
For highest accuracy, Zero AFTER changing the IF Bandwidth. Learn about
Zeroing in
How to set IF BW
— Press BW 2
— Then
— Then choose from the following:
Output Power
Set the power level out of the FieldFox to High, Low, or manually set power
level to a value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss
devices to place the signal farther from the noise floor. Increasing the power
can also help reduce trace noise. However, for devices that are sensitive to
high power levels such as amplifiers, use the Low power setting.
For best measurement accuracy, use the Manual power setting at -15 dBm.
After calibration, the power level can be decreased for amplifiers, or increased
for higher dynamic range.
Power Level settings in this mode will NOT change Power Level settings in
other modes. To help prevent damage to your DUT, use caution when
changing modes with your DUT connected to the FieldFox test ports.
Keysight N9938-90006 User's Guide
Resolution
1
0.0 (default)
2
0.00
"Zeroing" on page
IF BW
10 Hz | 100 Hz | 1 kHz | 10 kHz | 100 kHz
More 30 Hz | 300 Hz | 3 kHz | 30 kHz