Instrument Calibration
Over time, the active components in the FieldFox age and the performance
may degrade or drift. To ensure that the FieldFox is performing to the
published specifications, you should have an instrument calibration performed
periodically.
How Often Should Calibrate My Instrument?
You determine the calibration period which best meets your requirements.
However, a 12 month calibration cycle is appropriate for most users.
There are two things to consider: performance drift and connector wear.
The instrument specifications are set to consider the performance drift that
may occur over a 12 month period. Therefore, getting the instrument
calibrated at 12 month intervals ensures that the analyzer maintains
performance within the operating specifications. If you need the analyzer to
maintain more consistent operation, you may want to have the instrument
calibrated more often.
Connector wear is a bigger factor and depends on the number of connections
that are made. The test ports become noticeably worn after 500 to 700
connections. This could represent about 12 months with average use. With
more frequent connections, the calibration cycle should be sooner. You can
extend the time between calibrations and thereby save money by using
connector savers and by performing proper Connector Care. Learn more in
"Connector Care Review" on page
How do I Get an Instrument Calibration?
To get your Keysight instrument calibrated, send it to one of the Keysight
Technologies service centers. Please visit this website to learn how:
www.keysight.com/find/calibration.