Calibration for NA, CAT, and VVM Modes
measurement accuracy. Remember, CalReady was performed at the test ports.
Therefore, a CalReady calibration is most accurate when the DUT is directly
connected to the test ports. Learn more in
This setting does NOT survive Preset or Power ON/OFF.
— Press Cal 5 then
— Then press
To find the best choice for your DUT:
— Press Preset then Preset.
— Select the appropriate S-Parameter and other settings (frequency range,
— Press Trace 6 then
— Press
— Press Cal 5 then
— Select a CalReady Cal to compare with the current setting.
— Press Esc to exit the cal menu.
— View the differences in the two traces.
Keysight N9938-90006 User's Guide
More
CalReady
to toggle between the following selections:
—
2-Port Cal
(default) Corrects all four S-parameters. Requires a
forward and reverse sweep, which causes slower trace
measurements. Learn why in
—
Enh. Response
Corrects forward (S21 and S11) and reverse (S12 and
S22) measurements separately. Therefore, when measurements in
only one direction are required, this choice provides faster trace
measurements than a full 2-port cal. Also choose an Enhanced
Response Optimization. Learn more in
Optimization" on page
resolution, and so forth).
Math and Memory
Data &
Memory.
More
"CalReady" on page
"Calibration Type" on page
"Enhanced Response
141.
then Data->Mem.
127.
136.