Calibration for NA, CAT, and VVM Modes
measurement accuracy. Remember, CalReady was performed at the test ports.
Therefore, a CalReady calibration is most accurate when the DUT is directly
connected to the test ports. Learn more in
This setting does NOT survive Preset or Power ON/OFF.
— Press Cal 5 then
— Then press
To find the best choice for your DUT:
— Press Preset then Preset.
— Select the appropriate S-Parameter and other settings (frequency range,
— Press Trace 6 then
— Press
— Press Cal 5 then
— Select a CalReady Cal to compare with the current setting.
— Press Esc to exit the cal menu.
— View the differences in the two traces.
Keysight N9938-90006 User's Guide
More
CalReady
to toggle between the following selections:
2-Port Cal
(default) Corrects all four S-parameters. Requires a
forward and reverse sweep, which causes slower trace
measurements. Learn why in
Enh. Response
Corrects forward (S21 and S11) and reverse (S12 and
S22) measurements separately. Therefore, when measurements in
only one direction are required, this choice provides faster trace
measurements than a full 2-port cal. Also choose an Enhanced
Response Optimization. Learn more in
Optimization" on page
resolution, and so forth).
Math and Memory
Data &
Memory.
More
"CalReady" on page
"Calibration Type" on page
"Enhanced Response
141.
then Data->Mem.
127.
136.
143