SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
A color-coded legend for displayed traces is visible in the left pane of the SA
mode screen:
W = Clear/Write; M = MaxHold; m = MinHold; A = Average; V = View
How to display the Trace States
— Press Trace 6.
— Then
— Then
— Then choose from the following Trace States:
— Press
Trace 4 data WILL be overwritten by the FieldFox when using the
Independent Source Normalize feature
Generator " on page
corrections

Average Type

In SA Mode, there are four different processes in which Averaging is performed:
— Average Traces – Learn more in
— Detection Method Averaging – Learn more in
Keysight N9938-90006 User's Guide
Trace 1,2,3,4
repeatedly to select a trace number to display or change
its state. Traces are displayed in the following colors:
— Trace 1 – Yellow
— Trace 3 – Orange
State [current setting]
—Clr /Wr
(Clear/Write) Displays the actual measured data for each
sweep.
—MaxHold
Displays the maximum response of the input signal at each
data point over multiple sweeps. See also Average Count in
Count" on page
—MinHold
Displays the minimum response of the input signal at each
data point over multiple sweeps. See also Average Count in
Count" on page
—Average
Each data point is averaged over multiple sweeps as set by
the Average Count setting. Learn more in
page
200.
—View
Displays and then holds the last complete measurement trace on
the screen. This is similar to storing a trace in memory.
—Blank
Displays no trace.
Default All
to return all traces to their default settings: Trace 1
(Clear/Write); all other traces (Blank).
181) or using Field Strength antenna or cable
("Field Strength Measurements" on page
page
208.
Trace 2 – Blue
Trace 4 – Green
200.
200.
("Independent Source/Tracking
"Averaging" on page
"Average
"Average
"Average Count" on
161).
215.
"Detection Method" on
199