Pulse Measurements Mode – Option 330
Supported Power Sensors
IMPORTANT! When a LAN power sensor is initially connected to the LAN, it
may take ~10 seconds for the LAN power sensor to finish zeroing and for
the FieldFox to recognize the LAN power sensor. If your LAN power sensor
is not operating as expected:
Use an attenuator between the DUT and the power sensor when
measuring power levels that are higher than +20 dBm. The attenuator
value can be subtracted from the measurement using the Offset feature,
discussed in
A sequence of status lights are displayed on the base of the power sensor,
when the FieldFox is in Power Meter (LAN Sensor) mode.
— Green / Green blinking light indicates the power sensor's identification
— Amber light indicates Zeroing is being performed. And, blinks when a valid
— Red light indicates an error has occurred with the power sensor. Refer to
— No light indicates the FieldFox is not communicating with the power
572
Verify the latest firmware is loaded on the FieldFox. Refer to
www.keysight.com\find\fieldfoxsupport.
Verify the latest firmware is loaded on the Power Sensor. Refer to
https://www.keysight.com/main/editorial.
Press the LAN reset switch to reset the LAN sensor LAN configuration
to a known default state
Remove and re-connect the LAN power sensor
"Relative and Offset Power Measurements" on page
indicator is communicating. After successful identification, the green LED
stops blinking. No indication when the operating speed is 10 Mbps or
during line isolation.
link activity is detected.
power sensor documentation.
sensor. This is the case when in Hold mode or when NOT in Power Meter
Mode.
550.