Keysight Technologies B2900 Series User Manual

Keysight Technologies B2900 Series User Manual

Precision source/measure unit
Table of Contents

Advertisement

Quick Links

Keysight Technologies B2900 Series
Precision Source/Measure Unit
User's Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight Technologies B2900 Series

  • Page 1 Keysight Technologies B2900 Series Precision Source/Measure Unit User’s Guide...
  • Page 2 ERRORS OR FOR INCIDENTAL OR CONSE- in its End User License Agreement (EULA), a QUENTIAL DAMAGES IN CONNECTION WITH © Keysight Technologies 2011-2016 copy of which can be found at http:// THE FURNISHING, USE, OR PERFORMANCE www.keysight.com/find/sweula.
  • Page 4 COMPLIANCE WITH GERMAN NOISE REQUIREMENTS This is to declare that this product is in conformance with the German Regulation on Noise Declaration for Machines (Lärmangabe nach der Maschinenlärminformation-Verordnung -3.GSGV Deutschland). • Herstellerbescheinigung GERÄUSCHEMISSION Lpa < 70 dB am Arbeitsplatz normaler Betrieb nach DIN 45635 T.
  • Page 5 In addition, it violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for customer’s failure to comply with these requirements.
  • Page 6 Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel. Return the instrument to a Keysight Technologies sales or service office for services and repair to ensure that safety features are maintained.
  • Page 7 Safety Symbols The general definitions of safety symbols used on equipment or in manuals are listed below. Direct current. Alternating current. Earth ground terminal. Protective conductor terminal. For protection against electrical shock in case of a fault. Used with field wiring terminals to indicate the terminal which must be connected to ground before operating equipment.
  • Page 8 The RCM mark is a registered trademark of the Australian Communications Authority. This signifies compliance with the Australian EMC Framework Regulations under the terms of the Radio communications Act. This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada. This is the symbol for an Industrial, Scientific and Medical, Group 1 Class A product.
  • Page 9 Power Supply and Measurement Safety • Power Supply Safety This instrument can output high currents and voltages. Make sure that the load or device under test can safely handle the output current and voltage. Also, make sure that the connection leads can safely withstand the expected currents and are insulated for the expected voltages.
  • Page 10 • Source/Measure Terminals Source/measure unit, SMU, can simultaneously perform DC voltage or current output and measurement. Typical SMU has the Force, Sense, and Guard terminals as shown below. Normally the Force, Sense, and Guard are the same potential. Voltage marked around the terminals indicates the Protection Limits.
  • Page 11 High Vol tage Shock Hazard Keysight B2900 can force dangerous voltages (±210 V) at the High Force, Guard, and High Sense terminals. To prevent electric shock hazard, the following safety precautions must be observed during the use of Keysight B2900. •...
  • Page 12 Gefahr durch Hochspannung Von den Geräten Keysight B2900 können Spannungen an den Anschlüssen “High Force”, “Guard” und “High Sense” von bis zu 210 V ausgehen. Um elektrischem Schlag vorzubeugen, ist bei der Benützung der Geräte Keysight B2900 folgendes zu beachten. •...
  • Page 13 Danger de choc dû à une haute tension Une tension dangereuse (max. ± pour; 210 Vdc) émanant du dispositif Keysight B2900 peut être sortie aux bornes High Force, Guard et High Sense, d'appareil de protection ou de détection. Les précautions suivantes doivent être obserées contre commotion élec- trique accidentelle.
  • Page 14 高電圧感電注意 Keysight B2900 の High Force、Guard、High Sense 端子には、危険電圧が出力されること があります(最大 ±210 Vdc) 。感電事故防止のため、必ず以下の事柄を守ってくだ さい。 • 3 極電源ケーブルを使用して本器を接地してください。 • ドアを開くことによって開放されるインターロック回路を装備し、被測定デバ イスとのインタフェースを覆うことのできるシールド・ボックスを用意してく ださい。 • 測定を開始する前にはインターロック回路を本器の Interlock 端子に接続してく ださい。 • インターロック機能が正常であることを定期的に確認してください。 • High Force、Guard、High Sense 端子に繋がる接続部に触れる前には、本器の電源 を切断してください。また、測定系のキャパシタを放電してください。電源を 切らない場合は、以下の事項を全て実施してください。 • On/Off スイッチを押して On/Off スイッチが消灯したことを確認してくださ い。 •...
  • Page 15 Product Stewardship • Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/ electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE directive Annex 1, this product is classified as a “Monitoring and Control instrumentation”...
  • Page 16 In This Manual This manual describes the front panel operation, installation, and functions of the Keysight Technologies B2900 Precision Source/Measure Unit, SMU. This manual consists of the following chapters. "Getting Started" This chapter briefly explains how to use the Keysight B2900 by the front panel operation.
  • Page 17: Table Of Contents

    Keysight B2900 Series ........
  • Page 18 ............3-3 Inspecting the Shipment .
  • Page 19 Communicating Over the LAN ........3-31 Using the Graphical Web Interface ......3-31 Using Telnet .
  • Page 20 Result key group ..........4-40 Measure Result dialog box .
  • Page 21 To Perform Self-Test ........5-7 To Perform Self-Calibration .
  • Page 22 To Set Low Terminal State ........5-18 To Enable or Disable High Capacitance Mode .
  • Page 23 To Select a Program ........5-34 To Control Program Operation .
  • Page 24 Resources Used in the Expressions......6-18 Limit Test........... 6-20 Trace Buffer.
  • Page 25: Getting Started

    Getting Started...
  • Page 26 Getting Started This chapter describes basic operations for Keysight B2900. Before learning the details of B2900, let us briefly cover the use of the B2900 source/measure unit. The operations require only B2900 and the power cord. To get started with the operations, open the measurement terminals.
  • Page 27: Applying Dc Output

    Getting Started Applying DC Output Applying DC Output B2900 can be used as a DC voltage or current source. The following procedure configures B2900 as a voltage source and applies a voltage of +500 mV. Step 1. Setting the Source mode (voltage output) 1.
  • Page 28 Getting Started Applying DC Output This example sets the Source value to +500 mV. Step 3. Setting the Limit value (compliance current) 1. Press the assist key on a 1-ch model, or the assist key on a 2-ch Limit Ch1 Limit model.
  • Page 29: Performing Spot Measurement

    Getting Started Performing Spot Measurement Performing Spot Measurement B2900 can be used as a DC voltage or current meter. The following procedure performs one point current measurement. Step 1. Setting the measurement mode (current measurement) Press the assist key on a 1-ch model, or the assist key on a 2-ch Measure Ch1 Measure...
  • Page 30: Performing Sweep Measurement

    Getting Started Performing Sweep Measurement Performing Sweep Measurement B2900 supports sweep measurements. The following example applies a staircase sweep voltage and measures the current at each step voltage. Step 1. Press the key to display the Single view. Step 2. Set the Source mode, Limit value, and measurement mode as shown in “Applying DC Output”...
  • Page 31 Getting Started Performing Sweep Measurement 5. Use the rotary knob, arrow keys, or numeric/alpha keys to set the Start (sweep start value), Stop (sweep stop value), and Step (sweep step value) or Points (number of sweep steps). This example sets Start to 1 V, Stop to 2 V, and Points to 11. Step 4.
  • Page 32: Operation Tips

    Getting Started Operation Tips Operation Tips This section introduces key operation and status information on B2900. • “Changing the View Mode” • “Editing the Setup” • “Status Information” Figure 1-1 Changing the View Mode Single Graph B2901A B2902A Dual Single Ch1 Single Ch2 Graph B2911A...
  • Page 33 Getting Started Operation Tips Figure 1-2 Editing the Setup MOVE status EDIT status EDIT status MOVE status Select Press to select Press to fix For more information about the front panel keys, see “Hard Keys and Rotary Knob” on page 4-3.
  • Page 34: Operation Summary

    Getting Started Operation Summary Operation Summary This section summarizes front panel operations on B2900. • “Basic Operations” • “Channel Setup” • “Source Setup” • “Measurement Setup” • “Display Setup” • “File Operations” • “Miscellaneous Functions” • “Interface Setup” • “System Setup and Operations” For details about front panel operations, see Chapter 4, “Front Panel Reference.”...
  • Page 35 Getting Started Operation Summary Table 1-1 Basic Operations Task Relevant front panel key To turn the B2900 on/off Standby switch To change the display mode View To enable/disable the specified channel Ch 1 switch or Ch 2 switch On/Off On/Off To cancel the previous setup operation Cancel/Local To return to the upper menu of a function key or softkey...
  • Page 36 Getting Started Operation Summary Table 1-3 Source Setup Task Relevant front panel key To enable/disable source output Ch 1 switch or Ch 2 switch On/Off On/Off To select source output mode Mode, Ch1 Mode, or Ch2 Mode assist key To set source output value Source, Ch1 Source, or Ch2 Source assist key To set limit/compliance value Limit, Ch1 Limit, or Ch2 Limit assist key...
  • Page 37 Getting Started Operation Summary Task Relevant front panel key To set measurement ranging mode Speed assist key or Hide Sweep/Pulse/Trigger assist key on Single view To define automatic ranging operation Config > Measure > Ranging function keys To set measurement wait time Config >...
  • Page 38 Getting Started Operation Summary Table 1-7 Miscellaneous Functions Task Relevant front panel key To enable/disable Limit (Compliance) for Display > Pref. function keys each polarity To enable/disable Source/Limit real time Display > Pref. function keys update by using digit pointer and knob To see measurement result Result >...
  • Page 39 Getting Started Operation Summary Task Relevant front panel key To set LAN configuration I/O > LAN > Config function keys To see status of LAN interface I/O > LAN > Status function keys To reset all of LAN connections I/O > LAN > Reset function keys To sets LAN settings to the default settings I/O >...
  • Page 40 Getting Started Operation Summary Task Relevant front panel key To enable/disable beep and sound System > More > Sound function keys To select remote control command set System > More > Language function keys To enable/disable the function to restore Display >...
  • Page 41: Introduction

    Introduction...
  • Page 42 Introduction This chapter describes the basic functions and features of the Keysight B2900, and consists of the following sections. • “Keysight B2900 Series” • “Front View” • “Rear View” • “Source/Measure Unit” • “Operation and Functions” • “Software and Drivers”...
  • Page 43: Keysight B2900 Series

    Hence, B2900 can be used as a DC (constant) voltage/current source, sweep voltage/current source, pulse generator, arbitrary waveform generator, and multimeter. The product lineup of the Keysight B2900 series includes the following products. Table 2-1 Keysight B2900 Series...
  • Page 44: Front View

    Introduction Front View Front View This section describes the front view of the Keysight B2900 series. • Standby switch Turns the instrument on or off. When the power is on, the LED below the switch is turned green. • Display Displays the source setup, measurement result, status information, etc.
  • Page 45 Introduction Front View • Auto key Starts a repeat (continuous) measurement. If a repeat measurement is in progress, stops it. The repeat measurement is performed with the DC bias output of the Source value. And the measurement result is displayed on the Single view, Dual view, or Roll view.
  • Page 46 Introduction Front View • Numeric/alpha keys Used to enter the value of setup parameters such as the source output value, limit (compliance) value, and message, specified by the field pointer. • Rotary knob If the field pointer is in the MOVE (blue) status, turning the knob moves the pointer.
  • Page 47: Rear View

    Introduction Rear View Rear View This section describes the rear view of the Keysight B2900 series. • Channel 2 source/measure terminals Only on 2-channel models. Terminals for SMU channel 2. High Force, Low Force, High Sense, Low Sense, Guard, and chassis ground. For details, see “Connecting a DUT”...
  • Page 48 High Force, High Sense et Guard si la borne Interlock est fermée. NOTE Serial Number You will need the instrument’s serial number when using the Keysight Technologies telephone assistance program. The serial number label is attached to the bottom of the instrument. Keysight B2900 User’s Guide, Edition 5...
  • Page 49: Source/Measure Unit

    Introduction Source/Measure Unit Source/Measure Unit The source/measure unit (SMU) can apply voltage or current, and can measure current or voltage. Figure 2-1 shows a simplified SMU circuit diagrams. The SMU can perform the following operations. • Applies voltage and measures current •...
  • Page 50: Measurement Parameters

    Introduction Source/Measure Unit Measurement Parameters B2900 supports the following measurement parameters. • Current • Voltage • Resistance • Power Resistance data is given by Resistance=Vmeas/Imeas. Power data is given by Power=VmeasImeas. In the above formula, Vmeas is the voltage measurement data, and Imeas is the current measurement data.
  • Page 51: Output And Measurement Ranges

    Introduction Source/Measure Unit Output and Measurement Ranges This section describes typical specifications of the Keysight B2900 series. • Maximum voltage and current: see Figure 2-2. • Maximum power: 31.8 W • Output/measurement value and resolution: see Table 2-2 Table 2-7.
  • Page 52 Introduction Source/Measure Unit Table 2-2 Current Output Range Setting resolution Maximum voltage DC output current or Range pulse peak/base Pulse width t B2901A B2911A Pulsed value current B2902A B2912A output output 0  |I|  10.5 nA 210 V 210 V 50 s ...
  • Page 53 Introduction Source/Measure Unit Table 2-4 shows the limitations when using Channels 1 and 2 for DC output or Pulsed output with 50 s  t  t (=99.9999 ks). b. Maximum base current is 500 mA for the pulse with 50 s  t  1 ms, and 50 mA for the pulse with 50 s ...
  • Page 54 Introduction Source/Measure Unit Table 2-4 Limitations for using Channels 1 and 2 Channel 1 voltage V1 Channel 2 voltage V2 Current limit 0 < |V1|  6 V 0 < |V2|  6 V I1 + I2  4 A 6 V <...
  • Page 55 Introduction Source/Measure Unit Table 2-6 Voltage Measurement Range Range value Voltage measurement value Resolution 0  |V|  0.212 V 0.1 V 0.2 V 0  |V|  2.12 V 1 V 0  |V|  21.2 V 10 V 20 V 0 ...
  • Page 56: Operation And Functions

    Introduction Operation and Functions Operation and Functions Front Panel Interface 4.3” color TFT WQVGA (480272, with LED backlight) provides a graphical user interface (GUI) with the following features. • 4-view modes (Single, Dual, Graph, Roll view) • Easy-to-use key operations •...
  • Page 57 Introduction Operation and Functions Easy Access to Source/Measurement Capabilities • Settings: • V/I source mode • Source value • Limit (Compliance) value • Sweep parameters • Pulse parameters • Measure speed • Range setting • Trigger setting • Result Displays: •...
  • Page 58: Source/Measurement Capabilities

    Number of sweep steps: 1 to 100000. Or 1 to 2500 for the following B2900. • B2900 series without the license “SWS” • B2900 series with the firmware revision 3.0 or earlier • B2901A with the serial number MY51140470 or below •...
  • Page 59: Program And Interface Capabilities

    Introduction Operation and Functions • Time stamp • Measurement status • Source setting Maximum buffer size: 100000 points/channel Program and Interface Capabilities Programming Language The B2900 supports SCPI (Standard Commands for Programmable Instruments). • Default command set: Supports all of the advanced features of the B2900. •...
  • Page 60 Introduction Operation and Functions Digital I/O Interface • Connector: 25-pin female D • Input/output pins: 14 open drain I/O bits • Absolute maximum input voltage: 5.25 V • Absolute minimum input voltage: -0.25 V • Logic low maximum input voltage: 0.8 V •...
  • Page 61: Software And Drivers

    Introduction Software and Drivers Software and Drivers • Keysight B2900A Quick I/V Measurement Software The B2900 includes PC-based Keysight B2900A Quick I/V Measurement Software. This powerful software makes it easy to quickly setup and perform I/V measurements and to display the measurement data in a table or graph without the need to perform any programming.
  • Page 62: Accessories

    The Product Reference CD-ROM stores Keysight B2900A Quick I/V Measurement Software, sample programs, user manuals, and application notes. Available Accessories For the available accessories for the Keysight B2900 series, see the catalog Accessories for the Keysight B2900 Precision Instrument Family. To get the latest catalog, go to www.keysight.com/find/b2901a...
  • Page 63: Options

    Introduction Options Options Table 2-8 lists the options for the Keysight B2900. Table 2-8 Options Option Description Calibration ANSI Z540 compliant calibration Commercial calibration certificate with test data Printed Manual (User’s Guide) No printed manual (printed manual is not shipped) Traditional Chinese Simplified Chinese English...
  • Page 64 Introduction Options 2-24 Keysight B2900 User’s Guide, Edition 5...
  • Page 65: Installation

    Installation...
  • Page 66 Installation This chapter describes how to install Keysight B2900 and accessories and consists of the following sections. • “Inspecting the Shipment” • “Installing the Keysight B2900” • “Maintenance” • “Connecting a DUT” • “Installing the Interlock Circuit” • “Connecting to the Interfaces” •...
  • Page 67 Installation • Confirm that the switch does not turn red. On/Off • Open the fixture cover or the shielding box access door (open interlock). • Discharge any capacitors connected to a channel. • Warn persons working around the instrument about dangerous conditions. Une tension dangereuse (max.
  • Page 68: Inspecting The Shipment

    Installation Inspecting the Shipment Inspecting the Shipment Perform the following inspections when the Keysight B2900 and accessories arrive at your site. 1. Before unpacking any component, inspect all boxes for any signs of damage that might have occurred during shipment, such as: •...
  • Page 69: Checking For Errors

    Installation Inspecting the Shipment Figure 3-1 Display Example for 1-ch Model Figure 3-2 Display Example for 2-ch Model Checking for Errors Errors can be checked as described below. 1. Press the More > System > Error > Log function keys. This opens the Error Log dialog box.
  • Page 70: Installing The Keysight B2900

    Installation Installing the Keysight B2900 Installing the Keysight B2900 This section describes information you must pay attention to when installing the Keysight B2900. • “Safety Considerations” • “Environment” • “Connecting the Power Cord” • “Setting the Power Line Frequency” • “Setting Date and Time”...
  • Page 71: Connecting The Power Cord

    Installation Installing the Keysight B2900 Storage 10 % to 90 %RH, non-condensing • Altitude Operating 0 m to 2,000 m (6,561 ft.) Storage 0 m to 4,600 m (15,092 ft.) • Power requirement: 100-240 V (±10 %), 50/60 Hz • Maximum volt-amps (VA): 250 VA •...
  • Page 72 Installation Installing the Keysight B2900 Option 900 Option 901 Option 902 Option 903 • Plug: NEMA 5-15P, • Plug: IEC 60277-1, • Plug: AS/NZS 3112, • Plug: BS 1363/A, 125 V, 10 A 250 V, 10 A 250 V, 10 A 250 V, 10 A •...
  • Page 73: Setting The Power Line Frequency

    Installation Installing the Keysight B2900 Connect the power cord to the IEC 60320 connector on the rear of the instrument. If the wrong power cord was shipped with your instrument, contact Keysight Technologies. The AC input on the back of your instrument is a universal AC input. It accepts nominal line voltages in the range of 100 to 240 VAC.
  • Page 74: Rack Installation

    Installation Installing the Keysight B2900 Rack Installation NOTE Use the Rack Mount kit (Option 1CM) to rack mount the Keysight B2900. Keysight B2900 can be mounted in a 19-inch EIA rack cabinet. It is designed to fit in two rack-units (2U) of space. Remove the front and rear rubber bumpers and the handle before rack mounting the B2900.
  • Page 75: Maintenance

    Installation Maintenance Maintenance Maintenance should be performed periodically to keep the Keysight B2900 in good condition. If problems arise, contact your nearest Keysight Sales and Support Office. • “Cleaning” • “Self-test” • “Self-calibration” • “Calibration” Cleaning WARNING SHOCK HAZARD: To prevent electric shock, unplug the B2900 before cleaning.
  • Page 76: Self-Calibration

    Installation Maintenance • For preventive maintenance To perform The following procedure shows how to perform the self-test. self-test 1. If it is in the remote state, press the Cancel / Local key. 2. Press the On/Off switch and confirm that the switch turns off. 3.
  • Page 77: Calibration

    Installation Maintenance Calibration Calibration and adjustments must be performed periodically so that the instruments satisfy the specifications, and keep a good condition. It is recommended to perform the calibration once a year at least. For the calibration and adjustments, contact your nearest Keysight Sales and Support Office.
  • Page 78: Connecting A Dut

    Installation Connecting a DUT Connecting a DUT This section describes how to connect a device under test (DUT) to the Keysight B2900 source/measure terminals. NOTE Set the channel output off when changing the connections. If not, the DUT may be damaged.
  • Page 79: 2-Wire Connections Or 4-Wire Connections

    Installation Connecting a DUT Afin d'éviter de toucher l'extrémité du câble d'allongement ou l'aire de la borne avec l'appareil mis sous tension (AST), couvrez les conducteurs avec l'isolant. En outre, il est important de protéger la zone de la borne en utilisant le couvercle d'écran à...
  • Page 80: Floating

    Installation Connecting a DUT Figure 3-3 Dual View, 4-Wire Connection Setup for Channel 1 Floating With the default setting, Low Force terminal is connected to the chassis ground. However, it can be internally disconnected from the ground for making the floating condition.
  • Page 81 Installation Connecting a DUT Figure 3-4 Dual View, Floating Status for Channels 1 and 2 CAUTION Do not apply current to the chassis ground. Doing so will damage the B2900. WARNING To prevent electrical shock, do not touch any of measurement circuit at any time while a floating measurement is in progress.
  • Page 82: Using Test Leads

    Installation Connecting a DUT Using Test Leads B2900 source/measure terminals use banana jacks. For connecting a DUT, the following test leads are available. Figure 3-5 shows the connections for two-terminal device measurements. • Keysight U8201A Combo Test Lead Kit Couple of test leads, test probes, alligator clips, SMT grabbers, fine-tips test probes, and banana plugs, CAT III 1000 V, 15 A maximum Two kits are required for 4-wire connections.
  • Page 83: Using The N1295A Test Fixture

    Installation Connecting a DUT Using the N1295A Test Fixture Keysight N1295A is a test fixture which has four triaxial connectors and which supports two 2-wire connections. Requirements • Keysight N1295A test fixture furnished with pin clip wire 4 ea. and pin plug wire 2 ea.
  • Page 84: Using The 16442B Test Fixture

    Installation Connecting a DUT CAUTION Maximum voltage and current The test fixture and adapters must be used under the following limitations to prevent damage. Keysight N1294A-001: 250 V maximum, 42 V maximum for connecting N1295A Keysight N1294A-002: 250 V maximum, 42 V maximum for connecting N1295A Keysight N1295A: 42 V, 1.05 A maximum Using the 16442B Test Fixture...
  • Page 85 Installation Connecting a DUT Connections 1. Attach the banana to triaxial adapter to the B2900 source/measure terminals. See Figure 3-6. If there is some space as “Bad connection”, the contact is not enough. 2. Connect the triaxial cables between the adapter and the 16442B. Figure 3-8 shows a connection example.
  • Page 86 Installation Connecting a DUT NOTE Performing high voltage measurement When the Digital I/O interlock terminal is open, B2900 cannot apply high voltage over 42 V. To perform high voltage measurements, B2900 must be connected to the interlock circuit installed in 16442B. Prepare the N1294A-011 or 012 interlock cable, and connect it between the B2900 Digital I/O connector and the 16442B Intlk connector.
  • Page 87: Guarding

    Installation Connecting a DUT Guarding Guarding reduces the leakage current between the instrument and a DUT. This is important when you measure a low current. Figure 3-10 shows the theory of guarding. The buffer amplifier (1) keeps the potential of the Guard conductor at the same potential as the High Force conductor, so that the current does not flow between the High Force and Guard conductors.
  • Page 88: Installing The Interlock Circuit

    Installation Installing the Interlock Circuit Installing the Interlock Circuit The interlock circuit is a simple electric circuit, as shown in Figure 3-11. The circuit electrically opens when an access door is opened, and closes when the door is closed. B2900 cannot apply high voltages over 42 V when the Digital I/O interlock terminal is open.
  • Page 89 Installation Installing the Interlock Circuit • 25-pin D-sub connector (for wiring, mounted on the shielding box), 1 ea. • Connection wire • 25-pin D-sub cable, 1 ea. Procedure 1. Mount two mechanical switches onto your shielding box, so that the switches close when the access door is closed, and open when the door is opened.
  • Page 90 Installation Installing the Interlock Circuit Figure 3-12 Dimensions of the LED (Keysight part number 1450-0641) Units : mm Figure 3-13 Dimensions of the Interlock Switch (Keysight N1254A-402) 10.3 4.75 35.6 Switch off 15.3 10.3 15.9 18.8 C OM 22.2 27.8 Units: mm 3-26 Keysight B2900 User’s Guide, Edition 5...
  • Page 91: Connecting To The Interfaces

    Installation Connecting to the Interfaces Connecting to the Interfaces CAUTION Electrostatic discharges greater than 1 kV near the interface connectors may cause the unit to reset and require operator intervention. B2900 supports GPIB, LAN, and USB interfaces. All three interfaces are live at power-on.
  • Page 92 Installation Connecting to the Interfaces 1. If you have not already done so, install the Keysight IO Libraries Suite from the CD shipped with your product. 2. If you do not have a GPIB interface card installed on your computer, turn off your computer and install the GPIB card.
  • Page 93: Lan Interface

    Installation Connecting to the Interfaces You can also view the instrument’s VISA address from the front panel. Press the More function key, then the I/O > USB softkeys. The VISA address is shown in the USB Status dialog box. 4. You can now use the Interactive IO within Connection Expert to communicate with your instrument, or program your instrument using various programming environments.
  • Page 94 Installation Connecting to the Interfaces NOTE If you need to manually configure any instrument LAN settings, see “LAN Configuration dialog box” on page 4-48 for information about configuring the LAN settings from the front panel of the instrument. 3. Use the Connection Expert utility from Keysight IO Libraries Suite to add the B2900 and verify the connection.
  • Page 95: Communicating Over The Lan

    Installation Communicating Over the LAN Communicating Over the LAN Using the Graphical Web Interface Your Keysight B2900 Source/Measure Unit has a built-in graphical web interface that lets you control it directly from an Web browser on your computer. Multiple simultaneous connections are allowed, but only from one computer. With additional connections, performance will be reduced.
  • Page 96: Using Telnet

    Installation Communicating Over the LAN To launch the graphical web interface: 1. Open a Web browser on your computer. 2. Enter the instrument’s hostname or IP address into the browser’s Address field to launch the graphical web interface. The B2900 home page will appear. 3.
  • Page 97 Installation Communicating Over the LAN Once the port number is obtained, a control socket connection can be opened. As with the data socket, all commands to the control socket must be terminated with a newline, and all query responses returned on the control socket will be terminated with a newline.
  • Page 98: Using Digital I/O

    Installation Using Digital I/O Using Digital I/O B2900A has a Digital I/O connector, D-sub 25 pin female, for general purpose input/output (GPIO). It can be used for: • Trigger input • Trigger output • Digital signal input/output • Digital signal input •...
  • Page 99 Installation Using Digital I/O Figure 3-14 Digital I/O Internal Circuit to Input control to Digital I/O connector from Output control Vcc: +5 V R1: 5 kohm R2: 100 ohm Table 3-1 Digital I/O Pin Assignment Description Pin Number Description DIO 13 (bit 13) +5 V DIO 12 (bit 12) Interlock control...
  • Page 100: Accessory For Digital I/O Connector

    Installation Using Digital I/O a. Current limit: 600 mA (total current to the pins 22, 23, and 25) b. Used for the positive logic. Connected to the pin 25 for the negative logic. c. Used for the negative logic. Connected to the pin 17 for the positive logic.
  • Page 101: Front Panel Reference

    Front Panel Reference...
  • Page 102 Front Panel Reference This chapter provides the reference information of the Keysight B2900 front panel keys and display. • “Hard Keys and Rotary Knob” • “Display and Assist Keys” • “Function Keys” • “Config key group” • “Function key group” •...
  • Page 103: Hard Keys And Rotary Knob

    Front Panel Reference Hard Keys and Rotary Knob Hard Keys and Rotary Knob Standby switch Turns the instrument on or off. When the power is on, the LED below the switch is turned green. Trigger Starts a single (one shot) output/measurement or initiates trigger system. If a repeat measurement is in progress, stops it.
  • Page 104 Front Panel Reference Hard Keys and Rotary Knob Cancel / Local Cancels the setup operation if the instrument is in the local status. Returns the instrument to the local status if it is in the remote status. On/Off Used to enable or disable the SMU channel. Turns the channel off if it is in the output status even if it is in the remote status.
  • Page 105 Front Panel Reference Hard Keys and Rotary Knob Left and right keys If the field pointer is in the MOVE (blue) status, pressing the key moves the pointer. If the field pointer is in the EDIT (green) status, pressing the key changes the value of the setup parameter specified by the pointer.
  • Page 106: Display And Assist Keys

    Front Panel Reference Display and Assist Keys Display and Assist Keys Keysight B2900 provides several display modes which depend on the model as shown below. The display mode is changed by the View key. Pressing the key changes the mode as shown below. Single ...
  • Page 107: Dual View

    Front Panel Reference Display and Assist Keys Dual View Available on B2902A and B2912A. The upper half is for channel 1. The lower half is for channel 2. Each area displays the measurement results, source and measurement settings, and channel status. Most setup parameters can be edited on this display.
  • Page 108 Front Panel Reference Display and Assist Keys Assist keys In the following descriptions, Ch n indicates Ch 1 or Ch 2. Ch n Mode Changes the assist keys to VOLTS (V) and AMPS (I) used to select the source function. VOLTS (V) sets the voltage source.
  • Page 109: Single View

    Front Panel Reference Display and Assist Keys Single View Displays the measurement results, source and measurement settings, and channel status. Most setup parameters can be edited on this display. The lower half displays the setup parameters selected by Assist keys 2. Display area 1.
  • Page 110 Front Panel Reference Display and Assist Keys D2. Secondary measurement data. Turns orange if the data exceeds the Limit value. For the limit test, D2 displays PASS or FAIL instead of the measurement data. Assist keys 1 Mode Changes the assist keys to VOLTS (V) and AMPS (I) used to select the source function.
  • Page 111 Front Panel Reference Display and Assist Keys Assist keys 2 Speed Changes the assist keys to AUTO (1 PLC for 10 nA and 100 nA ranges, 0.01 PLC for other ranges), SHORT (0.01 PLC), MEDIUM (0.1 PLC), NORMAL (1 PLC), LONG (10 PLC), or MANUAL used to select the measurement speed.
  • Page 112 Front Panel Reference Display and Assist Keys Range parameters The following setup parameters are available. Source or Measure XXXX is decided by the source setting. Source Volts: Spot For the voltage source. Selects the range operation AUTO (automatic ranging) or FIXED (fixed range) for the constant voltage output and the source side voltage measurement.
  • Page 113 Front Panel Reference Display and Assist Keys For V/I, the channel performs measurement by using the present source/measure condition, and the resistance value is given by V/I calculation. Sweep parameters B2900 can be used as a sweep source, and supports the following sweep operations. The operation can be selected by using the assist keys displayed when the field pointer is in the EDIT (green) status on the Sweep Parameters field.
  • Page 114 Front Panel Reference Display and Assist Keys List sweep setup When the field pointer is in the EDIT (green) status on the LIST sweep Start/Stop/Points field, the following assist keys are available for setting the list sweep source. Edit Opens the List Sweep dialog box used to set the list sweep source.
  • Page 115 Front Panel Reference Display and Assist Keys Points Number of data points Maximum value Minimum value • Load List Sweep Data dialog box This dialog box provides the following fields for loading a list sweep data from a file stored on a USB memory. Highlight a file and press OK to load the data and close the dialog box.
  • Page 116 Front Panel Reference Display and Assist Keys Pulse parameters B2900 can be used as a pulse source, and supports pulsed output and measurement. Pulse ON or OFF can be selected by using the assist keys displayed when the field pointer is in the EDIT (green) status on the Pulse field. The following setup parameters are available.
  • Page 117 Front Panel Reference Display and Assist Keys Trigger parameters The B2900 supports the following trigger types for triggering source output and measurement. They are effective for setting triggers easily. The trigger type can be selected by using the assist keys displayed when the field pointer is in the EDIT (green) status on the Trigger field.
  • Page 118 Front Panel Reference Display and Assist Keys Table 4-1 Trigger Types and Setup Parameters Type Count Delay Period Trigger AUTO 0 second N.A. AUTO automatically SYNC entered value entered value N.A. AUTO TIMER entered value entered value entered value TIMER MANUAL entered value entered value...
  • Page 119: Graph View

    Front Panel Reference Display and Assist Keys Graph View Displays the graph for plotting the channel 1 and/or 2 measurement or math result. The Graph view can plot up to 5000 data. Display area 1. Graph display status ON or OFF. Only on 2-channel models. [n] is for channel n. 2.
  • Page 120 Front Panel Reference Display and Assist Keys Table 4-2 X and Y Axes Data Types for Graph View Data type Assist key Description I (A) AMPS (I) Current measurement data V (V) VOLTS (V) Voltage measurement data R () OHMS (R) Resistance measurement data P (W) WATTS (P)
  • Page 121: Roll View

    Front Panel Reference Display and Assist Keys NOTE If the number of data exceeds 5000 If the number of measurement data exceeds 5000, the Graph view and the Roll view plot the following data. where, n is integer, 1 to 5000. Number of data is 5001 to 10000: [2*(n-1)+1]th data Number of data is 10001 to 25000: [5*(n-1)+1]th data Number of data is 25001 to 50000: [10*(n-1)+1]th data...
  • Page 122 Front Panel Reference Display and Assist Keys Display area 1. Displays status ON or OFF, on B2911A Displays status Ch 1, Ch 2, or OFF, on B2912A Two lines can be displayed on the graph. [1] indicates the graph setup for line 1. [2] indicates the graph setup for line 2.
  • Page 123 Front Panel Reference Display and Assist Keys Assist keys Auto Scale Changes graph scale to fit the trace in the graph automatically. Dump Screen Opens the File Selection (Dump Screen) dialog box which is used to save the screen dump as a JPEG file. The file can be saved onto a USB memory connected to the front panel USB-A connector.
  • Page 124: Status Information

    Front Panel Reference Display and Assist Keys Status Information The status information is common for all display modes. It is displayed above the bottom softkeys associated with the function keys. Table 4-4 Status Indicator Label Color Description AUTO white Trigger auto. Automatic trigger is enabled now. white Trigger active.
  • Page 125: Function Keys

    Front Panel Reference Function Keys Function Keys Keysight B2900 has six function keys below the front panel display, and provides the following softkeys. Function keys 1 Config SMU configuration setup. Displays the softkeys for setting several functions of the SMU. See “Config key group”...
  • Page 126: Config Key Group

    Front Panel Reference Config key group Config key group Pressing the Config key displays the following softkeys for setting several functions of the SMU. Source Displays the following softkeys for setting the SMU source operation. Connection Channel operation and connection setup. See “Output Connection dialog box”...
  • Page 127: Output Filter Dialog Box

    Front Panel Reference Config key group Save, Recall Saves/recalls all setup information of the B2900. Five storage areas are available in the built-in non-volatile memory (NVRAM) for this purpose. Press the Save or Recall softkey to display the softkeys #1, #2, #3, #4, and #5 used for selecting the storage area.
  • Page 128: Output Connection Dialog Box

    Front Panel Reference Config key group Output Connection dialog box This dialog box provides the following parameters for setting the channel operation and connection. Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2) This field specifies the channel set by this dialog box.
  • Page 129: Sweep Dialog Box

    Front Panel Reference Config key group Sweep dialog box This dialog box provides the following parameters for setting details of the sweep source operation. Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2) This field specifies the channel set by this dialog box. Sweep Ranging Ranging mode used for the sweep source operation, BEST, AUTO, or FIXED.
  • Page 130: Ranging Dialog Box

    Front Panel Reference Config key group Ranging dialog box This dialog box provides the following parameters for setting details of the auto ranging operation for measurements. Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2) This field specifies the channel set by this dialog box. Current Auto Ranging Current measurement auto range operation, NORMAL, SPEED, or RESOLUTION (RESOLN).
  • Page 131: Wait Control Dialog Box

    Front Panel Reference Config key group Wait Control dialog box This dialog box provides the following parameters for setting the source and measurement wait time. The source wait time is defined as the time the source channel cannot change the output after the start of a DC output or the trailing edge of a pulse.
  • Page 132: Function Key Group

    Front Panel Reference Function key group Function key group Pressing the Function key displays the following softkeys for setting the math, limit test, and trace functions. Math Math function setup. See “Math Expression dialog box” on page 4-32. Limit Test Displays the following softkeys for setting the limit test function.
  • Page 133: Composite Limit Test Setup Dialog Box

    Front Panel Reference Function key group Composite Limit Test Setup dialog box This dialog box provides the following parameters for setting the limit test. Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2) This field specifies the channel set by this dialog box. Limit Test Composite limit test, ON or OFF Mode...
  • Page 134: Limit Test Setup Dialog Box

    Front Panel Reference Function key group For the DIO pin assignment, see “Using Digital I/O” on page 3-34. DIO pins assigned to the GPIO Pins, /BUSY, /SOT, or /EOT must be set to the DIGITAL I/O function by using the DIO Configuration dialog box.
  • Page 135 Front Panel Reference Function key group Fail on Only for compliance check. OUT or IN. Fail on=IN judges limit test failure if the channel goes into the compliance state. Fail on=OUT judges limit test failure if the channel comes out of the compliance state.
  • Page 136: Trace Buffer Setup Dialog Box

    Front Panel Reference Function key group Trace Buffer Setup dialog box This dialog box provides the following parameters for setting the trace function. The data specified by the Feed Data parameter can be stored in the trace buffer if the Buffer Control parameter is NEXT.
  • Page 137: Trigger Key Group

    Front Panel Reference Trigger key group Trigger key group Pressing the Trigger key displays the following softkeys for setting the trigger parameters in detail and controlling the trigger system. See Figure 6-8 on page 6-28. Config Opens a dialog box. See “Trigger Configuration dialog box”...
  • Page 138: Trigger Configuration Dialog Box

    Front Panel Reference Trigger key group Trigger Configuration dialog box This dialog box is used to set the trigger parameters in detail. Note that the overlapped parameter values are ignored and replaced with the settings made on the Single view. See “Trigger parameters”...
  • Page 139 Front Panel Reference Trigger key group Trigger Output Trigger output, ON or OFF Trigger output terminal, an internal bus (INT1, INT2), LAN port (LAN), or a digital I/O pin EXTn (n=1 to 14) If this parameter is ON, B2900 sends an output trigger when it changes the trigger state for the action specified by the Ch, Layer, and Action parameters.
  • Page 140: Result Key Group

    Front Panel Reference Result key group Result key group Pressing the Result key displays the following softkeys for displaying the measurement, limit test, and trace results. Measure Displays the measurement result. See “Measure Result dialog box” on page 4-40. Limit Test Displays the limit test result.
  • Page 141: Limit Test Result Dialog Box

    Front Panel Reference Result key group Max. Y-axis maximum value of the graph. Min. Y-axis minimum value of the graph. Limit Test Result dialog box This dialog box provides the following GUI for displaying the limit test result. The result data is shown in the data list area below the Length field. Only on 2-channel models.
  • Page 142: Trace Statistical Result Dialog Box

    Front Panel Reference Result key group Trace Statistical Result dialog box This dialog box provides the following GUI for displaying the trace statistical result. The result data is displayed in the Mean, Std. Dev., Min., and Max. fields. Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2) This field specifies the channel of the data to display.
  • Page 143: File Key Group

    Front Panel Reference File key group File key group Pressing the File key displays the following softkeys for saving a file on and loading a file from a USB memory connected to the front panel USB-A connector. Save Displays the following softkeys for saving a file. Pressing a softkey displays the File Selection dialog box.
  • Page 144: Program Key Group

    Front Panel Reference Program key group Program key group Pressing the Program key displays the following softkeys for setting and controlling the program memory. The program memory can be defined by using the SCPI commands while the B2900 is in the remote mode. Catalog Displays the Program Catalog dialog box which lists the programs saved in the program memory.
  • Page 145: I/O Key Group

    Front Panel Reference I/O key group I/O key group Pressing the I/O key displays the following softkeys for setting the I/O interfaces. Format Displays the softkeys for setting the data output format. See “Data Output Format” on page 4-46. Displays the following softkeys for managing the LAN interface. Config Displays the LAN Configuration dialog box used to set the configuration of the LAN interface.
  • Page 146: Data Output Format

    Front Panel Reference I/O key group Config Displays the DIO Configuration dialog box used to set the configuration of the Digital I/O interface. See “DIO Configuration dialog box” on page 4-48. Displays the DIO Read/Write dialog box used to read or write a value set to the Digital I/O interface.
  • Page 147: Format (Measure) Dialog Box

    Front Panel Reference I/O key group Format (Measure) dialog box This dialog box provides the following parameters for setting the elements of the measurement data output. Voltage Voltage data output, ON or OFF Current Current data output, ON or OFF Resistance Resistance data output, ON or OFF Source...
  • Page 148: Lan Configuration Dialog Box

    Front Panel Reference I/O key group ABSOLUTE Absolute value (ABS.). Returns the incremental value for the first timestamp data. DELTA Delta value (DELTA). Returns the incremental value for the previous timestamp data. LAN Configuration dialog box This dialog box provides the following parameters for setting the configuration of the LAN interface.
  • Page 149: Dio Read/Write Dialog Box

    Front Panel Reference I/O key group trigger input (TRIGGER IN), or high voltage state (HIGH VOLTAGE LAMP, HI-VOLT LAMP, only for pin 14) If Function is set to HIGH VOLTAGE LAMP, other setup parameters on this dialog box is ignored. Polarity Polarity of the input/output function, positive (POSITIVE, POS.) or negative (NEGATIVE, NEG.)
  • Page 150: Display Key Group

    Front Panel Reference Display key group Display key group Pressing the Display key displays the following softkeys for setting the display functions. The present setting is indicated by an asterisk on the softkey label. Remote Displays the following softkeys for enabling or disabling the front panel display when the B2900 is in the remote mode.
  • Page 151: Display Preference Dialog Box

    Front Panel Reference Display key group Display Preference dialog box This dialog box is used to set the following parameters. The settings are saved to the non-volatile memory. • V/I Limit for Each Polarity Selects the Limit (Compliance) setup operation. Individual setup ON Positive and negative limits can be set individually.
  • Page 152: System Key Group

    Front Panel Reference System key group System key group Pressing the System key displays the following softkeys used for several system setups. Error Displays the following softkeys to view or clear errors. Displays the Error Log dialog box which displays the SCPI errors.
  • Page 153 Front Panel Reference System key group Auto CLR Displays the following softkeys for setting the automatic clear of the timestamp. The present setting is indicated by an asterisk on the softkey label. Disables the automatic clear of the timestamp. Enables the automatic clear of the timestamp. Start-up Displays the System Start-up dialog box which provides the following setup parameters.
  • Page 154 Front Panel Reference System key group Info. Displays the following softkeys. Revision Displays the Revision dialog box which displays the model number, serial number, and firmware revision of the B2900. Date/Time Displays the Date and Time dialog box which is used to view or set the date and time.
  • Page 155: Front Panel Operations

    Front Panel Operations...
  • Page 156 Front Panel Operations This chapter explains how to use the Keysight B2900. • “Basic Operations” • “To Use Functions other than Measurement” • “To Set Source Output” • “To Execute Measurement” • “To Use Math Function” • “To Execute Limit Test” •...
  • Page 157: Basic Operations

    Front Panel Operations Basic Operations Basic Operations Use the field pointer to specify a setup field in the setup screen. The field pointer can take the following states. Change the status as needed during the setup of B2900. MOVE status Highlighted in blue.
  • Page 158: To Change The Settings On A Dialog Box

    Front Panel Operations Basic Operations When the pointer is on the decimal point, turning the knob moves the decimal point if the real time update (Immediate V/I Update by Knob) is OFF. NOTE Channel operation when using the rotary knob If the field pointer is in the EDIT (green) status on the Source or Limit (Compliance) field, turning the knob changes/updates the setting value of the source channel in real time.
  • Page 159: To Use Functions Other Than Measurement

    Front Panel Operations To Use Functions other than Measurement To Use Functions other than Measurement This section describes the following miscellaneous operations other than the source output and measurement functions. • “To Set Power Line Frequency” • “To Return to Factory Shipment Condition” •...
  • Page 160: To Set Power Line Frequency

    Front Panel Operations To Use Functions other than Measurement To Set Power Line Frequency 1. Press the More > System > PLC function keys. 2. Press 50 Hz or 60 Hz. To cancel the setting change, press the Cancel/Local key. To Return to Factory Shipment Condition 1.
  • Page 161: To Perform Self-Test

    Front Panel Operations To Use Functions other than Measurement To Perform Self-Test 1. Press the On/Off switch and confirm that the switch is turned off. 2. Disconnect test leads and cables from the channel terminals. 3. Press the More > System > Cal/Test > Self-Test function keys. The Confirmation dialog box opens.
  • Page 162: To Clear Error Buffer

    Front Panel Operations To Use Functions other than Measurement To Clear Error Buffer 1. Press the More > System > Error function keys. 2. To clear the error buffer, press Clear. To cancel the operation, press the Cancel/Local key. To Clear Timestamp 1.
  • Page 163: To Set Gpib Address

    Front Panel Operations To Use Functions other than Measurement NOTE License is required for each instrument (every serial number) to use the functions secured by license. To Set GPIB Address 1. Press the More > I/O function keys. 2. Press GPIB. The GPIB Configuration dialog box opens.
  • Page 164: To Save/Recall All Setup Information

    Front Panel Operations To Use Functions other than Measurement To Save/Recall All Setup Information This function saves/recalls all setup information of the B2900. For this function, five storage areas are available in the built-in non-volatile memory (NVRAM). To save the setup information, do the following steps. 1.
  • Page 165 Front Panel Operations To Use Functions other than Measurement To Enable/Disable Source/Limit Real Time Update by Using Digit Pointer and Knob For details on this function, see “Display Preference dialog box” on page 4-51. 1. Press the More > Display > Pref. function keys. The Display Preference dialog box opens.
  • Page 166: To Set Source Output

    Front Panel Operations To Set Source Output To Set Source Output This section describes the following source output operations. • “To Set Source Output Mode” • “To Apply DC Voltage/Current” • “To Stop Source Output” • “To Set Limit (Compliance) Value” •...
  • Page 167: To Set Source Output Mode

    Front Panel Operations To Set Source Output To Set Source Output Mode 1. For the Single view, press the Mode assist key. For the Dual view, press the Ch1 Mode or Ch2 Mode assist key. The field pointer moves to the Source mode setup field. 2.
  • Page 168: To Set Output Range

    Front Panel Operations To Set Source Output NOTE The Limit (Compliance) value can be set to both positive and negative directions individually by setting ON to the V/I Limit for Each Polarity field on the Display Preference dialog box. See “Display Preference dialog box”...
  • Page 169: To Set Sweep Output

    Front Panel Operations To Set Source Output 7. Press the rotary knob to change the pointer status to EDIT. 8. Enter the pulse peak value (Peak), delay time (Delay), or pulse width (Width). 9. Press the rotary knob or assist key to fix the setting value. 10.
  • Page 170: To Set List Sweep Output

    Front Panel Operations To Set Source Output NOTE Applying the sweep voltage/current Press the On/Off switch to start the Source value output. Press the Trigger key to perform the specified sweep output and measurement. To Set List Sweep Output For details on this function, see “List Sweep”...
  • Page 171: To Set Source Output Trigger Parameters

    Front Panel Operations To Set Source Output To Set Source Output Trigger Parameters For details on this function, see “Trigger System” on page 6-27 “Trigger parameters” on page 4-17. 1. Press the View key to display the Single view. 2. Press the Show Trigger assist key to display the Trigger parameters. 3.
  • Page 172: To Set Output Filter

    Front Panel Operations To Set Source Output To Set Output Filter For details on this function, see “Output Filter” on page 6-11 “Output Filter dialog box” on page 4-27. 1. Press the Config > Source > Filter function keys. The Output Filter dialog box opens. 2.
  • Page 173: To Enable Or Disable High Capacitance Mode

    Front Panel Operations To Set Source Output To cancel the setting change, press the Cancel/Local key. To Enable or Disable High Capacitance Mode For details on this function, see “High Capacitance Mode” on page 6-15. 1. Press the Config > Source > Connection function keys. The Output Connection dialog box opens.
  • Page 174: To Enable Or Disable Automatic Output-On Function

    Front Panel Operations To Set Source Output 4. To apply the settings, press Apply. To apply the settings and close the dialog box, press OK. To cancel the setting change, press the Cancel/Local key. To Enable or Disable Automatic Output-On Function For details on this function, see “Automatic Output-ON/OFF Function”...
  • Page 175: To Set Sweep Direction

    Front Panel Operations To Set Source Output The Sweep dialog box opens. 2. In the Ch field, set the channel to which the setting change is applied. 3. Set BEST, AUTO, or FIXED in the Sweep Ranging field. For details, see “Sweep dialog box”...
  • Page 176: To Execute Measurement

    Front Panel Operations To Execute Measurement To Execute Measurement This section describes the following measurement operations. • “To Enable Resistance Measurement” • “To Set Measurement Mode” • “To Perform Spot Measurement” • “To Stop Measurement” • “To Set Measurement Time” •...
  • Page 177: To Set Measurement Mode

    Front Panel Operations To Execute Measurement To Set Measurement Mode 1. For the Single view, press the Measure assist key. For the Dual view, press the Ch1 Measure or Ch2 Measure assist key. 2. Press AMPS (I) for the current measurement, VOLTS (V) for the voltage measurement, OHMS (R) for the resistance measurement, or WATTS (P) for the power measurement.
  • Page 178: To Set Measurement Range

    Front Panel Operations To Execute Measurement 4. When MANUAL is specified, the aperture time setup field and PLC setup field are displayed. Enter and fix a value in either of the fields to set the measurement time. To Set Measurement Range For details on this function, see “Ranging Mode”...
  • Page 179: To Set Measurement Trigger Parameters

    Front Panel Operations To Execute Measurement The channel starts the specified voltage/current output. While the switch is turned green, the channel continues the output and changes the value immediately if it is changed. 8. Press the Trigger key. The sweep measurement is performed and the measurement result is displayed in a graph.
  • Page 180: To Set Measurement Auto Range Operation

    Front Panel Operations To Execute Measurement The Wait Control dialog box opens. 2. Set each parameter in the Measure column. 3. To apply the settings, press Apply. To apply the settings and close the dialog box, press OK. To cancel the setting change, press the Cancel/Local key. To Set Measurement Auto Range Operation For details on this function, see “Ranging dialog box”...
  • Page 181: To Use Math Function

    Front Panel Operations To Use Math Function To Use Math Function This section describes how to use the math function. For details on this function, see “Math Expression” on page 6-17. Also, for details on the setup parameters, see “Math Expression dialog box” on page 4-32.
  • Page 182: To Execute Limit Test

    Front Panel Operations To Execute Limit Test To Execute Limit Test This section describes how to set the limit test and how to display the result of the limit test. For details on this function, see “Limit Test” on page 6-20.
  • Page 183: To Set Individual Limit Tests

    Front Panel Operations To Execute Limit Test 12. In the /SOT field, set the DIO pin number used for the start-of-test (SOT) signal output. 13. In the /EOT field, set the DIO pin number used for the end-of-test (EOT) signal output.
  • Page 184 Front Panel Operations To Execute Limit Test OUT: Failure if the channel comes out of the compliance state. • In the Fail Pattern field, set a bit pattern that represents the limit test fail state (Fail). 9. When the test mode is LIMIT and the operation mode is GRADING, configure the following settings: •...
  • Page 185: To Display Limit Test Result

    Front Panel Operations To Execute Limit Test To Display Limit Test Result For details on the setup parameters, see “Limit Test Result dialog box” on page 4-41. 1. Press the Result > Limit Test function keys. The Limit Test Result dialog box opens. 2.
  • Page 186: To Use Trace Buffer

    Front Panel Operations To Use Trace Buffer To Use Trace Buffer This section describes how to set the trace buffer and how to display the statistical data. For details on this function, see “Trace Buffer” on page 6-23. To use the trace buffer, set the trace buffer and source output and measurement conditions, and then press the Trigger key.
  • Page 187: To Display Statistical Data

    Front Panel Operations To Use Trace Buffer To Display Statistical Data For details on the setup parameters, see “Trace Statistical Result dialog box” on page 4-42. 1. Press the Result > Trace function keys. The Trace Statistical Result dialog box opens. 2.
  • Page 188: To Use Program Memory

    Front Panel Operations To Use Program Memory To Use Program Memory This section describes how to select and run a memory program. For details on this function, see “Program Memory” on page 6-25. See also “Program key group” on page 4-44.
  • Page 189: Function Details

    Function Details...
  • Page 190 Function Details This chapter describes the following functions of the Keysight B2900. • “Limit/Compliance” • “Ranging Mode” • “Measurement Time” • “Pulse Output” • “Sweep Output” • “List Sweep” • “Output Filter” • “Over Voltage/Current Protection” • “Output-Off Status” • “Automatic Output-ON/OFF Function”...
  • Page 191: Limit/Compliance

    Function Details Limit/Compliance Limit/Compliance Limit/Compliance is the output limiter for preventing damage to the test device from overcurrent or overvoltage. Voltage compliance is for current output channels, and current compliance is for voltage output channels. When a channel reaches compliance, the channel acts as a constant voltage source or a constant current source.
  • Page 192: Ranging Mode

    Function Details Ranging Mode Ranging Mode The following ranging modes are available for performing source output or measurement. • FIXED (fixed range) The channel uses the specified range only. • AUTO (auto range) The channel automatically uses the range which provides the best resolution for the source output value or the measurement value.
  • Page 193: Measurement Time

    Function Details Measurement Time Measurement Time Measurement time depends on aperture time, measurement range, and other measurement conditions, and can be expressed by the following formula: Measurement time = Aperture time + Overhead time Aperture time is the time required for measurement, and does not include such factors as range change or data compensation, which would be the overhead time.
  • Page 194 Function Details Measurement Time 2. Measure delay The measurement delay time is defined as the time from trigger to start of a measurement. 3. Source wait The source wait time is defined as the time the source channel cannot change the output value after starting an output.
  • Page 195: Pulse Output

    Function Details Pulse Output Pulse Output The source/measure unit SMU can apply the pulsed voltage or current. Figure 6-2 shows a pulsed sweep output example. For a pulsed bias output, focus the shape of a pulse only. To Control Pulse Output/Measure Timing The pulse output and measurement timing can be controlled by the following parameters.
  • Page 196: To Set Pulse Output

    Function Details Pulse Output Figure 6-2 Pulse Output and Measurement Timing, Sweep Output Example Voltage or current Trigger Trigger Trigger Stop value Source delay (1) Measure delay (2) Pulse delay (3) Source wait (4) Measure wait (5) Start value Previous value : Measurement Pulse width (6) Time...
  • Page 197 Function Details Sweep Output Sweep Output The source/measure unit SMU can apply a sweep voltage or current. It supports several shapes of sweep output, as shown in Figure 6-3. The SMU not only performs the sweep output but performs the measurement for each sweep step as shown in Figures and 6-2.
  • Page 198: List Sweep

    Function Details List Sweep List Sweep The list sweep function is effective for performing an arbitrary waveform output. The source/measure unit (SMU) can apply the waveform as shown in Figure 6-4, and measure voltage or current at each output value. The source output and the measurement can be performed in the following minimum interval.
  • Page 199: Output Filter

    Function Details Output Filter Output Filter The filter is mounted on the source/measure unit (SMU). It assures clean source output without spikes, noise, and overshooting. However, using a filter may increase the SMU settling time. To set the filter, see “Output Filter dialog box”...
  • Page 200: Over Voltage/Current Protection

    Function Details Over Voltage/Current Protection Over Voltage/Current Protection The over voltage/current protection is effective for preventing damage to the test device from overcurrent or overvoltage. If this function is enabled, the source/measure unit (SMU) sets the output to 0 V and sets the output switch to off automatically and immediately when it reaches the compliance status.
  • Page 201: Output-Off Status

    Function Details Output-Off Status Output-Off Status The output-off status is the condition which is set on the source/measure unit (SMU) automatically, immediately after its output is turned off. The status must be specified before enabling the source output. The available conditions are shown in Table 6-1 To set the output-off status, see “Output Connection dialog box”...
  • Page 202: Automatic Output-On/Off Function

    Function Details Automatic Output-ON/OFF Function Automatic Output-ON/OFF Function The automatic output-ON/OFF function defines the source channel’s output ON/OFF operation when the trigger system changes the status. • Automatic-ON function If this function is enabled, the source/measure unit (SMU) automatically turns the channel output ON just before the trigger system is initiated by an SCPI command, not by a front panel operation.
  • Page 203: High Capacitance Mode

    Function Details High Capacitance Mode High Capacitance Mode High capacitance mode is effective for the measurement of a capacitive load greater than 0.01 F. If the measurement result data is not stable, set this function ON. The measurement data may come stable. This function is effective for measurements of capacitive devices up to 50 F.
  • Page 204: Resistance Measurement

    Function Details Resistance Measurement Resistance Measurement B2900 supports resistance measurement. If the measurement parameter is set to resistance OHMS (R), the source/measure unit (SMU) automatically sets the current source and voltage measurement operation to perform resistance measurement. For performing accurate measurement, B2900 provides the compensation function. NOTE To Enable Resistance Measurement The resistance measurement operation must be specified.
  • Page 205: Math Expression

    Function Details Math Expression Math Expression B2900 provides a math function for performing calculations using the measurement result data. The calculation result can be displayed and used for the limit test and trace statistics. For the predefined math expressions, see “Predefined Math Expressions”.
  • Page 206: Resources Used In The Expressions

    Function Details Math Expression This function is effective for reducing measurement errors in low resistance measurements. VARALPHA Calculates varistor alpha using the following formula. VARALPHA = log(CURR[c][1] / CURR[c][0]) / log(VOLT[c][1] / VOLT[c][0]) where, CURR[c][0] and VOLT[c][0] are the measurement data at a point on a varistor’s non-linear I-V characteristics curve, and CURR[c][1] and VOLT[c][1] are the data at the another point.
  • Page 207 Function Details Math Expression Table 6-2 Reserved Variables Reserved variable Description Scalar Vector SOUR[c] SOUR[c][] Source output setting data VOLT[c] VOLT[c][] Voltage measurement data CURR[c] CURR[c][] Current measurement data RES[c] RES[c][] Resistance measurement data TIME[c] TIME[c][] Time (timestamp) data a. The numeric suffix [c] is effective for specifying the channel. For exam- ple, use CURR2 to read the current spot measurement data of the chan- nel 2.
  • Page 208: Limit Test

    Function Details Limit Test Limit Test Limit test is a pass/fail judgement performed for a measurement data or math result data obtained by a channel. It can be performed if both individual limit test and composite limit test are set to ON. Maximum of twelve limit tests can be defined and used for the bins of composite limit test.
  • Page 209 Function Details Limit Test Figure 6-5 Composite Limit Test Flowchart Example for Grading Mode Display example of PASS status when Zoom is ON: START Single view Dual view Wait for SOT Source and Sense Perform Pass ? Display “FAIL” Limit X ? Last Limit ? Immediate ? Pass ?
  • Page 210 Function Details Limit Test Figure 6-6 Composite Limit Test Flowchart Example for Sorting Mode Display example of PASS status START when Zoom is ON: Single view Wait for SOT Source and Sense Dual view Perform Limit X ? Pass ? Last Limit ? Display “FAIL”...
  • Page 211: Trace Buffer

    Function Details Trace Buffer Trace Buffer The trace buffer collects the test result data until a buffer full is detected. The maximum data size is 100,000 blocks for each channel. The data flow is shown in Figure 6-7. One data block may contain multiple data, such as voltage measurement data, current measurement data, resistance measurement data, source output data, calculation result data, limit test data, time data, and status data.
  • Page 212 Function Details Trace Buffer NOTE Do not forget to save or read the data before turning the B2900 off. The trace buffer is cleared by turning the instrument off. Figure 6-7 Data Flow to Trace Buffer for Each Channel Sense (Voltage, Current) VOLT CURR...
  • Page 213: Program Memory

    If frequently used command strings are stored in the program memory, interface/computer activity is minimized. The program memory can be defined by using SCPI commands while B2900 is in the remote mode. See Keysight B2900 series SCPI Command Reference. • Number of programs saved in the memory: 100 •...
  • Page 214: Channel Grouping

    Function Details Channel Grouping Channel Grouping This section applies only to 2-channel models. This function is used to control the channel output timing automatically so that the channel keeps the output while the other channel performs measurement. The grouped channels start the source output in the order of channel number, then start the measurement at the same time, and keep the output until the measurement is completed.
  • Page 215: Trigger System

    Function Details Trigger System Trigger System B2900 supports the ARM-TRIGGER model described in 1999 SCPI Command Reference. Operation summary of this trigger model is described below. Also see Figure 6-8. The trigger system can be initiated by pressing the Trigger key on the front panel. For the trigger setup parameters, see “Trigger parameters”...
  • Page 216 Function Details Trigger System Figure 6-8 B2900 Trigger System Trigger key Auto key Idle Initiated ? INT1-2 INT1-2 Trigger output Trigger output ARM Layer EXT1-14 EXT1-14 before ARM after ARM ARM count = 0 AUTO Event TIMER Last count ? INT 1-2 event detector...
  • Page 217: Trigger Source

    Function Details Trigger System Trigger Source B2900 supports the following trigger sources. The trigger source must be set to the ARM event and the TRIGGER event individually. • AUTO (automatic internal, AINT): Trigger source best suited for the present operating mode is automatically selected by the internal algorithms. •...
  • Page 218 Function Details Trigger System • Measurement wait time control is set to OFF. • Measurement ranging mode is set to the fixed mode. • To synchronize measurement: • Trigger source is set to the same mode. • Delay time is set to the same value. •...
  • Page 219: Trigger Output

    Function Details Trigger System Trigger Output B2900 can output trigger at the timing of Trigger output shown in Figure 6-8. The trigger output terminal can be selected from the following terminals. • Internal bus, INT1, INT2 • LAN port • Digital I/O pin, EXT1, EXT2, EXT3, EXT4, EXT5, EXT6, EXT7, EXT8, EXT9, EXT10, EXT11, EXT12, EXT13, EXT14 The trigger output can be set by using...
  • Page 220: File Access Function (Easy File Access)

    Function Details File Access Function (Easy File Access) File Access Function (Easy File Access) File access function uses MTP (media transfer protocol) to easily download B2900 files (measurement data, instrument settings, and screen images) to your PC. NOTE This function is available for the B2900 firmware revision 3.0 or later. And the “SWS”...
  • Page 221: Interlock Function

    Function Details Interlock Function Interlock Function The interlock function is designed to prevent electrical shock when a user touches the measurement terminals. If the interlock terminal is open, maximum output is limited to 42 V. To perform high voltage measurement greater than 42 V, connect the interlock terminal to an interlock circuit of a test fixture or a shielding box.
  • Page 222: Over Temperature Protection

    Function Details Over Temperature Protection Over Temperature Protection The over temperature protection is effective for preventing damage to a channel from over temperature. If B2900 is used in an environment over 30 C, channel outputs are limited to the value less than the maximum value. If a channel output reaches the limit, all channels automatically and immediately set the output to 0 V and set the output switch to off.
  • Page 223: Initial Settings

    Function Details Initial Settings Initial Settings The B2900 is initialized by turning the B2900 on, the *RST command, or the device clear. Initial settings of the B2900 are shown in this section. Table 6-6 System Initial Settings Setup item Power on Reset ...
  • Page 224 Function Details Initial Settings Setup item Power on Reset Program variables None Not changed  Program selection None  Program status Idle Table 6-7 SMU Initial Settings Setup item Power on Reset Low terminal state GROUNDED Not changed  Output state ...
  • Page 225 Function Details Initial Settings Setup item Power on Reset  Voltage sweep points  Voltage sweep start  Voltage sweep stop  Voltage list points  Voltage list values 100 A  Current source 100 A  Current protection level ...
  • Page 226 Function Details Initial Settings Setup item Power on Reset  Settling time  Measure function Current measure  Auto aperture  Aperture time 0.1 PLC  Voltage measure range mode AUTO  Voltage measure range limit 0.2 V  Voltage measure range ...
  • Page 227 Function Details Initial Settings Table 6-8 Trigger System Initial Settings Setup item Power on Reset  ARM count  ARM source AINT 100 s  ARM timer  ARM delay time  ARM bypass  Trigger count  Trigger source AINT 10 s for B291x ...
  • Page 228 Function Details Initial Settings Table 6-9 LXI Trigger Events Initial Settings Setup item Power on Reset  Event domain  LAN event “WaitingForAcquir eArm1”, “WaitingF orTransitionArm1” , “WaitingForAcqu ireTrigger1”, “Wait ingForTransitionTr igger1”, “Measurin g1”, “Settling1” “WaitingForAcquir eArm2”, “WaitingF orTransitionArm2” , “WaitingForAcqu ireTrigger2”, “Wait ingForTransitionTr igger2”, “Measurin...
  • Page 229 Function Details Initial Settings Table 6-10 Calculate Functions Initial Settings Setup item Power on Reset  Composite limit test result transfer timing  Composite limit test result automatic clearance  Composite limit test fail bit pattern All 0  Composite limit test pass bit pattern All 0 ...
  • Page 230 Function Details Initial Settings Setup item Power on Reset  Math function catalog “POWER”, “OFFCOMPOHM ”, “VOLTCOEF”, “VARALPHA”  Math function unit name “W”  Offset value for test  Offset value enable  Trace feed source SENSe  Trace control NEVer ...
  • Page 231 Function Details Initial Settings Table 6-11 Non-volatile Communication Settings Setup item Factory default setting DHCP Enabled IP address 169.254.5.2 Subnet mask 255.255.0.0 Default gateway 0.0.0.0 Obtain DNS server from DHCP Enabled DNS server 0.0.0.0 WINS server 0.0.0.0 Hostname K-B29xxA-nnnnn B29xxA: model number Desired hostname nnnnn: suffix of serial number.
  • Page 232 Function Details Initial Settings Setup item Factory default setting Web interface Enabled Command prompt for a Telnet session B2900A> Welcome message for a Telnet session Welcome to Keysight B2900A Series Table 6-12 Other Non-volatile Settings Setup item Factory default setting Channel grouping “1”...
  • Page 234 This information is subject to change without notice. © Keysight Technologies 2011-2016 Edition 5, December 2016 *B2910-90010* B2910-90010 www.keysight.com...

Table of Contents