Section 7: TSP command reference
trigger.timer[N].passthrough
This attribute enables or disables the timer trigger pass-through mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
passthrough = trigger.timer[N].passthrough
trigger.timer[N].passthrough = passthrough
passthrough
N
Details
When pass-through mode is enabled, triggers are passed through immediately and initiate the delay.
When disabled, a trigger only initiates a delay.
Example
trigger.timer[1].passthrough = true
Enables pass-through mode on trigger timer 1.
Also see
trigger.timer[N].reset()
7-416
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Instrument reset
Recall setup
Trigger timer N reset
The state of pass-through mode; set to one of the following values:
▪
true: Enabled
▪
false: Disabled
Trigger timer number (1 to 8)
(on page 7-417)
Where saved
Default value
Not saved
false (disabled)
2651A-901-01 Rev. C October 2021