Section 7: TSP command reference
localnode.autolinefreq
This attribute enables or disables automatic power line frequency detection at start-up.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
flag = localnode.autolinefreq
localnode.autolinefreq = flag
flag
Details
When this attribute is set to true, the power line frequency is detected automatically the next time
the Model 2651A powers up. After the power line frequency is automatically detected at power-up,
the localnode.linefreq attribute is set automatically to 50 or 60.
If the localnode.linefreq attribute is explicitly set, localnode.autolinefreq is automatically
set to false.
When using this command from a remote node, replace localnode with the node reference, for
example node[5].autolinefreq.
Also see
localnode.linefreq
7-172
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Not applicable
The auto line frequency detection setting:
true: Enable automatic line frequency detection at start-up
false: Disable automatic line frequency detection at start-up
(on page 7-174)
Where saved
Default value
Nonvolatile memory
true (enabled)