Section 7: TSP command reference
smuX.cal.unlock()
This function enables the commands that change calibration settings.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.cal.unlock(password)
password
Details
This function enables the calibration functions to change the calibration settings.
The password when the instrument is shipped from the factory is "KI0026XX".
Example
smua.cal.unlock("KI0026XX")
Also see
Adjustment
(on page 12-15)
smuX.cal.lock()
smuX.cal.password
smuX.cal.restore()
smuX.cal.state
smuX.contact.calibratehi()
This function adjusts the high/sense high contact check measurement.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.contact.calibratehi(cp1Measured, cp1Reference, cp2Measured, cp2Reference)
cp1Measured
cp1Reference
cp2Measured
cp2Reference
Details
Contact check measurement calibration does not require range information.
Typically, points one and two are near 0 Ω and 50 Ω, respectively.
All four measurements (cp1Measured, cp1Reference, cp2Measured, and cp2Reference) must
be made with the calibration set that is active. If not, corruption of the calibration constants may
result.
7-236
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Calibration password
(on page 7-230)
(on page 7-231)
(on page 7-233)
(on page 7-235)
Affected by
The value measured by this SMU for point 1
The reference measurement for point 1 as measured externally
The value measured by this SMU for point 2
The reference measurement for point 2 as measured externally
Where saved
Unlocks calibration.
Where saved
2651A-901-01 Rev. C October 2021
Default value
Default value