Model 2651A High Power System SourceMeter® Instrument Reference Manual
smuX.measure.filter.enable
This command enables or disables filtered measurements.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
filterState = smuX.measure.filter.enable
smuX.measure.filter.enable = filterState
filterState
Details
This command enables or disables the filter.
Example
smua.measure.filter.count = 10
smua.measure.filter.type = smua.FILTER_MOVING_AVG
smua.measure.filter.enable = smua.FILTER_ON
Sets the filter count to 10.
Sets the filter type to moving average.
Enables the filter.
Also see
Filters
(on page 3-3)
reset()
(on page 7-197)
setup.recall()
smuX.measure.filter.count
smuX.measure.filter.type
smuX.reset()
(on page 7-264)
2651A-901-01 Rev. C October 2021
Affected by
Instrument reset
SMU reset
Recall setup
The filter status; set to one of the following values:
0 or smuX.FILTER_OFF: Disables the filter
1 or smuX.FILTER_ON: Enables the filter
(on page 7-221)
(on page 7-250)
(on page 7-252)
Section 7: TSP command reference
Where saved
Default value
Saved setup
0 (smuX.FILTER_OFF)
7-251