Section 7: TSP command reference
smuX.source.offlimitY
This attribute sets the limit (current or voltage) used when the source-measure unit (SMU) is in normal output-off
mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
sourceLimit = smuX.source.offlimitY
smuX.source.offlimitY = sourceLimit
sourceLimit
Y
Details
Setting the current limit to lower than 1 mA may interfere with operation of the contact check function.
See smuX.contact.check() and smuX.contact.r() for details.
Example
smua.source.offlimiti = 10e-3
Also see
smuX.contact.check()
smuX.contact.r()
smuX.source.offfunc
smuX.source.offmode
7-276
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
SMU reset
Instrument reset
Recall setup
Set to the limit to be used when the SMU is in normal output-off mode
SMU source function (v = voltage, i = current)
(on page 7-239)
(on page 7-240)
(on page 7-275)
(on page 7-277)
Where saved
Default value
Saved setup
Current: 1e-3 (1 mA)
Voltage: 40 (40 V)
Changes the normal output-off mode limit to
10 mA.