Section 7: TSP command reference
serial.read()
This function reads available characters (data) from the serial port.
Type
TSP-Link accessible
Function
Yes
Usage
data = serial.read(maxchars)
data
maxchars
Details
This function reads available characters from the serial port. It does not wait for new characters to
arrive. As long as maxchars is less than 200 characters, all characters that are received by the serial
port (before the serial.read() command is executed) are returned. If too many characters are
received between calls to this function, the RS-232 buffers overflow and some characters may be
lost.
Call this function as many times as necessary to receive the required number of characters. For
optimal performance, use a small delay between repeated calls to this function.
The data returned is the raw data stream read from the port. No characters, such as control
characters or terminator characters, are interpreted.
If you attempt to use this function when the serial port is enabled as a command interface, a settings
conflict error is generated.
Example
data = serial.read(200)
print(data)
Also see
serial.write()
(on page 7-217)
7-216
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Where saved
A string that consists of all data read from the serial port
An integer that specifies the maximum number of characters to read
Read data from the serial port.
Output:
John Doe
The above output indicates that the string "John Doe" was read
from the serial port.