Section 2: General operation
To edit compliance limit values:
1. Press the LIMIT key.
2. Select the type of compliance (CURRENT, VOLTAGE, or POWER).
3. Press the navigation wheel or the ENTER key to enter edit mode. The EDIT indicator is
illuminated.
4. Change the compliance value (see
5. When finished, press the ENTER key (the EDIT indicator is not illuminated).
The up and down range keys change the format of the limit value.

Beeper

The Model 2651A includes a beeper. When it is enabled, a beep indicates one of the following
actions have occurred:
A front-panel key was pressed: A short beep, similar to a key click, is issued.
The navigation wheel was turned or pressed: A short beep is issued.
The output source was changed: A longer beep is issued when you select the OUTPUT
ON/OFF control (turn the output on or off).
To turn the beeper on or off from the front panel:
1. Press the MENU key, and then select BEEPER.
2. Select one of the following:
ENABLE
DISABLE
To turn the beeper on or off from the TSP command interface:
Set the beeper.enable attribute. For example, to enable the beeper, send:
beeper.enable = 1
2-18
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Setting a value
(on page 2-16)).
2651A-901-01 Rev. C October 2021