Section 7: TSP command reference
smuX.measure.filter.count
This command sets the number of measured readings that are required to yield one filtered measurement.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
filterCount = smuX.measure.filter.count
smuX.measure.filter.count = filterCount
filterCount
Details
This attribute sets the size of the stack used for filtered measurements.
Example
smua.measure.filter.count = 10
smua.measure.filter.type = smua.FILTER_MOVING_AVG
smua.measure.filter.enable = smua.FILTER_ON
Sets the filter count to 10.
Sets the filter type to moving average.
Enables the filter.
Also see
Filters
(on page 3-3)
reset()
(on page 7-197)
setup.recall()
smuX.measure.filter.enable
smuX.measure.filter.type
smuX.reset()
(on page 7-264)
7-250
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Instrument reset
SMU reset
Recall setup
The number of readings required for each filtered measurement (1 to 100)
(on page 7-221)
(on page 7-251)
(on page 7-252)
Where saved
Default value
Saved setup
1
2651A-901-01 Rev. C October 2021