Section 7: TSP command reference
trigger.timer[N].count
This attribute sets the number of events to generate each time the timer generates a trigger event.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
count = trigger.timer[N].count
trigger.timer[N].count = count
count
N
Details
If the count is set to a number greater than 1, the timer automatically starts the next trigger timer
delay at the expiration of the previous delay.
Set the count to zero (0) to cause the timer to generate trigger events indefinitely.
If you use the trigger timer with a trigger model, make sure the count value is the same or more than
any count values expected in the trigger model.
Example
print(trigger.timer[1].count)
Read trigger count for timer number 1.
Also see
trigger.timer[N].clear()
trigger.timer[N].delay
trigger.timer[N].reset()
trigger.timer[N].delay
This attribute sets and reads the timer delay.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
interval = trigger.timer[N].delay
trigger.timer[N].delay = interval
interval
N
7-412
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Instrument reset
Recall setup
Trigger timer N reset
Number of times to repeat the trigger (0 to 1,048,575)
Trigger timer number (1 to 8)
(on page 7-411)
(on page 7-412)
(on page 7-417)
Affected by
Instrument reset
Recall setup
Trigger timer N reset
Delay interval in seconds (8 µs to 100 ks)
Trigger timer number (1 to 8)
Where saved
Default value
Not saved
1
Where saved
Default value
Not saved
10e-6 (10 µs)