Model 2651A High Power System SourceMeter® Instrument Reference Manual
TSP-Net compared to TSP-Link to communicate with TSP-enabled
devices
The TSP-Link
communication occurs between the Model 2651A and another TSP-enabled instrument.
One of the advantages of using the TSP-Link network interface is that TSP-Link connections have
three trigger lines that are available to each device on the TSP-Link network. You can use any one of
the trigger lines to perform hardware triggering between devices on the TSP-Link network. Refer to
Hardware trigger modes
However, if the distance between the Model 2651A and the TSP-enabled device is longer than
15 feet, use TSP-Net commands.
TSP-Net instrument commands: General device control
The following instrument commands provide general device control:
tspnet.clear()
tspnet.connect()
tspnet.disconnect()
tspnet.execute()
tspnet.idn()
(on page 7-446)
tspnet.read()
(on page 7-447)
tspnet.readavailable()
tspnet.reset()
tspnet.termination()
tspnet.timeout
tspnet.write()
TSP-Net instrument commands: TSP-enabled device control
The following instrument commands provide TSP-enabled device control:
tspnet.tsp.abort()
tspnet.tsp.abortonconnect
tspnet.tsp.rbtablecopy()
tspnet.tsp.runscript()
2651A-901-01 Rev. C October 2021
®
network interface is the preferred communication method for most applications where
(on page 3-65) for details.
(on page 7-442)
(on page 7-443)
(on page 7-444)
(on page 7-445)
(on page 7-448)
(on page 7-449)
(on page 7-449)
(on page 7-450)
(on page 7-455)
(on page 7-451)
(on page 7-452)
(on page 7-453)
(on page 7-454)