Installation

Section 6
Space allotment for a still should include a 6" clearance at the sides and
top to allow for cover removal and air circulation. Unit must be placed
within 4 feet of water supply and should not be mounted over electrical
equipment to avoid shock hazard (see warning below).

Wall Mounting

The MEGA-PURE 11A water still has slotted holes for wall mounting. To
support the weight of this unit, use 1/4" x 1 1/4" lag bolts for mounting.
An optional wall mounting bracket is available for the Automatic
Collection System. Demineralizers have slotted holes for wall mounting.
Demineralizer location is not critical, though it should be close enough to
the still to make all necessary connections and should not be mounted over
electrical equipment to avoid shock hazard (see warning).
Warning Do not mount your MEGAPURE 11A water still directly over
equipment that requires electrical service. Routine maintenance of this unit
may involve water spillage and subsequent electrical shock hazard if
improperly located.
Warning Wall composition and construction, as well as fastener type, must
be considered when mounting this unit. The mounting surface and
fasteners selected must be capable of supporting a minimum of 125 lbs.;
inadequate support and/or fasteners may result in damage to mounting
surface and/or equipment. If you are unsure of mounting surface
composition, condition and construction, or correct fasteners, consult your
building maintenance group or contractor.

Bench Mounting

No special consideration for a "still only" setup. When used with the ACS,
the MEGA-PURE 11A water still must be placed on a 6" platform to
allow for gravity feed of distilled water to the ACS.